6.0 PROGRAMMING
The Plum XL Micro/Macro has the following delivery
modes:
o
Primary only delivery
o
Secondary only delivery
o
Piggyback delivery
When a rate and a VTBI are entered for the primary line
and no settings are entered for the secondary line, the
Plum XL will deliver primary only. Likewise, when a rate
and a VTBI are entered for the secondary line and no
settings are entered for the primary line, the Plum XL
delivers secondary only. When a rate and a VTBI are
entered for the primary and secondary lines (piggyback
delivery), the Plum XL completes secondary delivery
before it begins primary delivery.
The Plum XL retains all previous therapy settings
and fluid delivery data in its memory until the settings
are cleared by the user. Check the primary and
secondary settings during the initial setup to confirm
that all settings are correct. Confirm the proper clearing
of the total volume delivered before use.