38
Plum
XL
Micro/Macro
Use 19-gauge or larger needle or catheter for viscous
fluids if operating at rates greater than 500 mL/hr.Use
a cassette with a capped secondary port when delivering
viscous fluids on the secondary line.
8.6
Backpriming
Backpriming is not recommended for reconstituting
secondary containers containing dry powders.
To avoid pressurization when backpriming into a syringe
or a vial, the user must ensure that these containers
have sufficient empty space to accept the backprimed
fluid.
8.7
General
Possible explosion hazard exists if used in the presence
of flammable anesthetics.
Product damage may occur unless proper care is
exercised during unpacking and installation. Do not use
the Plum XL if it appears damaged in any way.
Do not place Plum XL in service if it fails the self-test (see
Section 3.2, Self-Test
for detailed information).
Use of radio frequency emitting devices such as
electrosurgical units, cellular telephones and two-way
radios in close proximity of this device may affect its
operation.
The screen displays VTBI in 0.1-mL increments from
0.1 to 99.9 mL. 100 to 9999 mL are displayed in 1-mL
increments. Any fraction of a milliliter delivered is not
displayed, but is retained in memory.
Keep the cassette door securely closed while the Plum
XL is not in use, to avoid cassette door damage.