Error Messages and Troubleshooting
66
The response is slow
The indicated value does not change
The measured value is out of the measurement range
When the measured value is below the display range, “Ur” appears. When the measured value
is over the display range, “Or” appears.
Cause
How to solve problem
The electrode is dirty.
Clean the electrode.
The electrode is broken.
Replace the electrode.
The response of electrode is affected
by the sample solution.
It is important to select an electrode that is
appropriate for the sample. Consult your dealer. To
confirm an electrode that is appropriate for the
sample, check the pH electrode selection guide in
our catalogue, or refer to our website.
Cause
How to solve problem
The electrode is cracked.
Replace the electrode.
The electrode is not connected
correctly.
Connect the electrode correctly.
The instrument is in HOLD state.
Cancel the HOLD state.
Instrument defect.
Consult your dealer.
Cause
How to solve problem
Sample is out of the measurement
range.
Use a sample within the measurement range.
Electrode is not immersed enough to
cover liquid junction.
The electrode must be immersed up to the liquid
junction.
As a guide, immerse to at least 3 cm from the tip of
the electrode.
The electrode cable is broken.
Replace the electrode.
Calibration is not performed or
performed incorrectly.
Perform calibration correctly.
Instrument defect.
Check as explained below.
Содержание LAQUA-PC1500
Страница 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 24: ...Product Overview 6...
Страница 46: ...pH Setup 28...
Страница 54: ...Conductivity Setup 36...
Страница 58: ...TDS Setup 40...
Страница 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Страница 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Страница 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Страница 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Страница 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Страница 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Страница 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Страница 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Страница 91: ...Printout Summary 73 Resistivity Conductivity...
Страница 92: ...Printout Summary 74 Salinity TDS...
Страница 93: ...Printout Summary 75 Printer Format Data Log...
Страница 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Страница 99: ......