
ORP Calibration
13
● ORP/mV Calibration
Calibration is necessary for accurate ORP measurement. To perform ORP calibration, follow
the procedure detailed below:
Prerequisites
• Clean the ORP electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the meter and plug in the ORP electrode.
• Prepare standard solution required for calibration.
• Ensure that the meter is in mV measurement mode.
• Dip the ORP electrode into the standard solution ensuring that the solution level is at least
3 cm from the electrode tip.
• Absolute value measurement mode and relative value measurement mode are the two
types of measurement mode available for ORP (mV) measurement.
• In absolute value measurement mode, the handheld meter displays the actual voltage
value.
• In relative value measurement mode, user can adjust the absolute mV value by calibration.
If the mV value is adjusted, the meter automatically indicates relative mV value as
RmV
.
The adjustment mV is applied as an offset to the absolute mV value.
• In the relative mV mode, the absolute mV value can be adjusted by ± 200 mV.
To abort an ongoing calibration process at any point of time, press the
key.
Note
Tip
Содержание LAQUA-PC1500
Страница 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 24: ...Product Overview 6...
Страница 46: ...pH Setup 28...
Страница 54: ...Conductivity Setup 36...
Страница 58: ...TDS Setup 40...
Страница 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Страница 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Страница 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Страница 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Страница 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Страница 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Страница 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Страница 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Страница 91: ...Printout Summary 73 Resistivity Conductivity...
Страница 92: ...Printout Summary 74 Salinity TDS...
Страница 93: ...Printout Summary 75 Printer Format Data Log...
Страница 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Страница 99: ......