Temperature Calibration
22
Calibration
1. After placing the electrode in the solution, press the
key.
2. Press the
key to switch to temperature calibration mode. Meter displays measured
temperature value.
3. Use the
keys to adjust the temperature to the required value.
4. Press the
ENT
key to save calibration data.
5. Meter displays
DONE
indicating end of the temperature calibration procedure.
key
key
key
key
Содержание LAQUA-PC1500
Страница 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 24: ...Product Overview 6...
Страница 46: ...pH Setup 28...
Страница 54: ...Conductivity Setup 36...
Страница 58: ...TDS Setup 40...
Страница 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Страница 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Страница 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Страница 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Страница 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Страница 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Страница 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Страница 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Страница 91: ...Printout Summary 73 Resistivity Conductivity...
Страница 92: ...Printout Summary 74 Salinity TDS...
Страница 93: ...Printout Summary 75 Printer Format Data Log...
Страница 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Страница 99: ......