pH Calibration
11
■ Calibration
This section describes the calibration procedures using the LAQUA-PC1500 benchtop meter,
pH and conductivity electrodes.
● pH Calibration
Calibration is necessary for accurate pH measurement. To perform pH calibration, follow the
procedure detailed below:
Prerequisites
• Clean the pH electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the meter and plug in the pH electrode.
• Prepare the buffers required for calibration.
• Keep the meter in pH measurement mode.
• Dip the pH electrode at least 3 cm in the buffer.
• Perform two-point calibration using:
pH 7 and 4 for acidic sample.
pH 7 and 10 for alkaline sample.
• Perform three-point calibration using pH 7, 4 and 10 if you are unsure of the expected
sample pH value. It is recommended to calibrate with pH 7 first.
• Default buffer setup is
BUFF USA
. If you want to change to
BUFF NIST
or
BUFF DIN
,
refer to “P1.1 Buffer Setup” on page 26.
• To abort an ongoing calibration process at any point of time, press the
key.
• It is recommended to clear the previous calibration data before performing calibration. For
erasing the calibration data, refer to “P1.2 Erase Calibration Data” on page 27.
Note
Tip
Содержание LAQUA-PC1500
Страница 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 24: ...Product Overview 6...
Страница 46: ...pH Setup 28...
Страница 54: ...Conductivity Setup 36...
Страница 58: ...TDS Setup 40...
Страница 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Страница 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Страница 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Страница 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Страница 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Страница 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Страница 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Страница 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Страница 91: ...Printout Summary 73 Resistivity Conductivity...
Страница 92: ...Printout Summary 74 Salinity TDS...
Страница 93: ...Printout Summary 75 Printer Format Data Log...
Страница 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Страница 99: ......