49
8 Troubleshooting
USB is not
displayed
though the PC
is connected on
the memory
display.
PC is not
started up.
Start up the PC.
Connection
failure or
malfunction of
the micro USB
cable
Insert the micro USB
cable again. If the
problem is not
solved, replace the
cable.
Use of a
charge-only
type cable
Use a micro USB
cable that enables
data transfer.
Data is not
transferred
although the
[SET/
]
button is
pressed on the
memory display.
The input
screen of
application
software of PC
is not active.
Click (or tap) the
input screen of
application software
to activate it.
After the [SET/
] button is
pressed on the
memory display,
[USb] is
displayed at the
glossiness
display position.
Connection
failure or
malfunction of
the micro USB
cable
Insert the micro USB
cable again. If the
problem is not
solved, replace the
cable.
Use of a
charge-only
type cable
Use a micro USB
cable that enables
data transfer.
Problem
Cause
Action to take
Ref.