4 How to use
Calibration
Calibration is recommended in the following cases.
• When the surrounding environment has drastically changed
(example: the product is brought from the inside the room to
the outside)
• When an hour or more has elapsed after measurement is
started or after calibration
1.
Attach the protective cover to the IG-340 main unit.
Be careful not to cause looseness and backlash of the
cover when attaching.
2.
Press and hold the [CAL] button for two seconds.
Calibration starts.
[CAL] blinks at the glossiness display position during
calibration. When the calibration is complete, [CAL] turns
off.
Fig. 11 Calibration execution
The IG-340 main unit is adjusted in such a way that the
measurement value obtained when the standard plate has
been measured is set as a standard value (refer to
"Standard value setting" (page 39)).