Chapter 6 Maintenance and Troubleshooting
6.2 Troubleshooting
70
●
The temperature display is blinking.
The temperature display does not change from 25ºC.
The temperature measurement exceeds the measurement range.
Measurement range: -10 – 100.0ºC
●
Measurements are not repeatable
●
When the printer will not print even though it is connected.
Check the following points:
・
Is the printer turned ON?
・
Has a printer error occurred?
・
Is there printing paper? Has the paper jammed?
・
When running a test print according to the manual, does it print out
correctly?
Cause
How to solve problem
The temperature of
the sample solution
exceeds the
measurement range.
Check the temperature of the sample solution
and change to a sample solution that has a
temperature within the measurement range.
The thermistor
connection within the
electrode is severed
or shorted.
Measure the resistance of the temperature
sensor connector. If it is 50 k
Ω
or more at
room temperature, replace the electrode.
The electrode
connector is not
attached properly.
Attach the electrode connector properly, so
that the O-ring on the temperature connector
disappears from sight.
The main unit of the
meter may be
defective.
In Temperature Display Calibration mode ( "
Temperature zero adjustment [item No.
02]" page 34), check whether or not the
"Minus" display appears, regardless of
whether or not there is a temperature connec-
tor.
There is a problem
with the setting for
the temperature
display calibration
mode (see page 34).
Initialization of setting [item No. 07]"
Cause
How to solve problem
Some effects of the
sample solution
The pH or other properties of the sample
solution may have changed over time, making
repeatability poor.