Chapter 6 Maintenance and Troubleshooting
6.2 Troubleshooting
66
●
ERR 03 Electrode stability error
Explanation
The electric potential did not stabilize within three minutes.
Cause
How to solve problem
This is caused by the
sample solution (when
the sample solution is
pure water or another
solution with low
conductivity or the
concentration or
temperature change).
Press the MEAS key again while “HOLD” is
either blinking or steadily lit in the display, to
measure the sample using instantaneous
value
measurement.
The electrode is dirty.
Wash the electrode.
The temperature of the
sample solution is
fluctuating.
Measure after the sample solution
temperature stabilizes.