Chapter 6 Maintenance and Troubleshooting
6.2 Troubleshooting
DS-51
69
●
The response is slow
●
The indicated value does not change, or there is absolutely no response
●
The measured value is blinking
The measured conductivity value exceeds
the
measurement
parameters (when conductivity value is displayed).
Display range: 0.00 – 19.99 (when cell constant is 100 m
-1
)
Check this point
As shown in the diagram, use a jumper wire or bent paper clip to short
the meter by touching both the center pin and some metal part in the
electrode connector.
If the flaching measured value disappears when this done, the meter is
nomal.
Cause
How to solve problem
Some effects of the
sample.
Response time may slow down, depending on
the properties of the sample solution.
Cause
How to solve problem
The key-lock function
is ON.
Turn the power OFF, and then turn it back ON
again.
The system is
locked.
Turn the power OFF, and then turn it back ON
again.
The electrode
connector is not
attached correctly.
Attach the electrode connector correctly.
meter is defective.
Contact your local HORIBA distributor.
Cause
How to solve problem
The sample solution
is inappropriate.
Change to a sample solution with properties
within the measurement range.
The electrode cable
has been severed.
Replace the electrode.
The main body of the
meter is defective.
Check the point described below.