Chapter 6 Maintenance and Troubleshooting
6.2 Troubleshooting
DS-51
67
●
ERR 09 Printer error
Explanation
If a problem occurs with the printer, turn OFF the power to the meter,
perform the appropriate measure below, and turn the power to the meter
back ON.
●
ERR 10 Data memory over
Explanation
The number of data items has exceeded the limit of the memory.
●
ERR 11 Cell constant out of range
Explanation
The cell constant is out of the range of 0.7 to 1.3.
Delete data stored in the memory after confirming their contents.
Cause
How to solve problem
The printer paper is
jammed.
Remove the jammed paper.
There is no printer
paper.
Load the printer with paper.
There is a problem
with the printer
connection.
Reconnect the printer after making sure there
is nothing wrong with the connector parts.
The printer is
defective.
Replace the printer.
Cause
How to solve problem
Memory ove
Delete data stored in the memory after
confirming their contents.
Cause
How to solve problem
COND electrode is at
the end of its useful life.
Replace the electrode.
Improper standard
solution
Prepare new standard solution.