Chapter 3 Functions
3.3 Setting modes
34
●
Temperature zero adjustment [item No. 02]
This mode uses a known temperature to calibrate the temperature
compensation value.
1.
Immerse the electrode in a liquid with a known temperature, and
set the temperature using the
▲
and
▼
keys.
Setting range: 0.0 to 100.0 °C
The temperature sensor attached to the electrode maintains an accuracy of
±
1
°
C, even without calibration. The mode should be used when a greater
precision than
±
1
°
C is required.
2.
Use the MODE key to proceed to Automatic power-off setting
(item No. 03).