![horiba DS-51 Скачать руководство пользователя страница 49](http://html.mh-extra.com/html/horiba/ds-51/ds-51_instruction-manual_2162205049.webp)
Chapter 4 RS-232C communications
4.2 Command list
40
4.2
Command list
Use <CR><LF> as the terminator for serial communication commands.
All the commands (except the ON-LINE/OFF-LINE command) are valid
only in the ON-LINE mode. (An error message is returned in the OFF-
LINE mode.)
The meter returns a response to any operation made in the following
format:
OK<CR><LF>
If the pH meter does not accept the operation, it returns an error mes-
sage in the following format:
ER,n<CR><LF>
n= 0: Communication error
1: When a non-existent command is input.
2: When a timing command is input to which the meter cannot
respond.
3: When the numerical value in the command is out of the setting
range.