Chapter 6 Maintenance and Troubleshooting
6.1 Conductivity electrode maintenance
DS-51
63
Chapter 6
Maintenance and Troubleshooting
This chapter explains how to perform daily meter maintenance and how
to deal with error messages.
Daily maintenance is vital in assuring accurate measurement and
preventing breakdowns before they occur. Maintenance of the
electrodes is especially important; if ignored, various problems and
erroneous measurements may result. This COND meter is equipped
with a convenient error message function. If an error message is
displayed, be sure to take appropriate action.
6.1
Conductivity electrode maintenance
Refer to the electrode operation manuals for how to maintain each
electrode.
Long-term storage
When an electrode will not be used for a long period of time, store it
after performing the following procedure.
Also, perform maintenance on the electrode every three to six months.
1.
Remove the electrode from the COND meter.
2.
Use pure (de-ionized) water to wash away any sample solution
that may have adhered to the electrode.
3.
Wash the inside of the electrode protective cap with pure (ion
exchange) water, then, after shaking out the water, fill the cap with
enough pure (de-ionized) water to soak the sponge.
4.
Place the electrode protective cap on the electrode.