HORIBA
Security
NOx Converter Checker CC-100E
7/ 24
Electrical safety
DANGER
Electrical voltage
Danger of electric shock
There are parts in the unit that are under dangerous voltage
a)
The unit may only be opened by instructed, skilled personnel.
b)
Disconnect the unit from the mains at all poles when carrying out maintenance
work on electrical components. To do this, pull out the mains plug. Simply
switching off the unit at the mains switch is not sufficient!
c)
Secure the unit against unintentional restart.
d)
Work on the opened device when it is live may only be carried out by a specialist
who is familiar with the associated dangers.
Switch off in an emergency
WARNING
Emergency
Uncontrollable operation of the appliance, occurrence of fire and smoke.
a)
Switch off the unit and disconnect the mains plug.
b)
Close the valves on the gas cylinders or the external feeds.
Hazardous substances
WARNING
Toxic gases
The unit handles toxic gases. Uncontrolled leakage of these gases can cause a
serious accident such as poisoning.
a)
Connect the system according to this manual.
b)
Use the specified connections as well as hoses and pipes and connect them
gas-tight.
c)
Check that the applied gas pressure corresponds to the nominal value.
d)
Ensure that your working environment is properly ventilated.
CAUTION
Handling auxiliary materials and consumables
Auxiliary materials and consumables may contain substances that are
hazardous to health or the environment.
a)
Follow the instructions for action written in the safety data sheets of the
auxiliary materials and consumables.
b)
Wear your personal protective equipment if necessary.