HORIBA
Operation
NOx Converter Checker CC-100E
17/ 24
Set ozone concentration and start ozone generator
Tap the "OGU" symbol. The window for setting the ozone concentration appears
Fig. 8:
Ozone generation
Set the "counts" using the 100s / 10s / 1s key and the "Up" and "Down" keys. Switch the generator on with
the "On" key.
You can then work through the test sequence with the "Up" and "Down" buttons.
The set "counts" are displayed in the main screen.
NOTE
Relationship between counts and NO concentration
The value for "counts" should be 10 times the NO concentration in ppm.
Example:
For an NO concentration of 100 ppm, "counts" must be set to 1000
Supply external gas
To supply N2 (SV2) or NO (SV3) externally, open the respective inlet valve as described above.
Fig. 9:
Opening the valves
NOTE
Supply external gas
The inlet valves SV2 and SV3 are interlocked. Only either N2 or NO can be supplied.