Maintenance
HORIBA
20/ 24
NOx Converter Checker CC-100E
Situation
Possible causes
Remedial action
–
The pressure generated at the
outlet is out of specification.
Regulate the pressure to a value in
the specified range.
The reaction speed is slow.
–
The residual pressure of the
gas is too low.
Regulate the pressure to a value in
the specified range. Connect a new
bottle if necessary.
–
The inlet pressure is too low.
Regulate the pressure to a value in
the specified range.
–
A leak has occurred.
Check the gas lines and screw
connections. Connect the gas lines
properly.
–
The flow is too low
Ensure a sufficient flow rate.
6.4 Spare parts
The following spare parts can be obtained from the manufacturer:
Spare part
Type
Item no.
Inlet valve 1 (oxygen)
Type 6011
1000067176
Inlet valve 2 (nitrogen)
Type 127
1000067178
Inlet valve 3 (nitrogen oxide)
Type 127
1000067178
Pressure regulator 1
UA13-S202-AV
3011014227
Pressure regulator 2
UA13-L202-SV
3011021069
Capillary 1
SA-101
3014056923
Capillary 2
SA-102-J
3014056400
Ozone generator
OGU
3014061759