Operation Manual HORIBA APDA-371
Particulate Monitor
Date:
April, 2010
______________________________________________________________________________________________________________________
______________________________________________________________________________________________________________________
HORIBA Europe GmbH, Julius-Kronenberg-Str. 9, D-42799 Leichlingen, Telefon: +49(0)2175-8978-0, Fax: +49(0)2175-8978-50
Page 3
C
H
A
N
X
ID
POWER
COM
SIG
APDA-371 Back Panel Sensor Input Terminal
500 Series Sensor Wiring Connections for APDA-371
BX-590 Wind Direction Sensor
BX-591 Wind Speed Sensor
Terminal Block
Cable Wire Color
Terminal Block
Cable Wire Color
SIG
Yellow
SIG
Yellow
COM
Black/Shield
COM
Black/Shield
POWER
Red
POWER
Red
ID
Green
ID
Green
BX-592 Ambient Temp Sensor
BX-593 Relative Humidity Sensor
Terminal Block
Cable Wire Color
Terminal Block
Cable Wire Color
SIG
Yellow
SIG
Yellow
COM
Black/Shield
COM
Green/Shield
POWER
Red
POWER
White
ID
Green
ID
Red
BX-594 Barometric Pressure Sensor
BX-595 Solar Radiation Sensor
Terminal Block
Cable Wire Color
Terminal Block
Cable Wire Color
SIG
White
SIG
Yellow
COM
Black/Shield
COM
Black/Shield
POWER
Red
POWER
Red
ID
Yellow
ID
Green
BX-596 Temperature/Pressure Combo
Sensor
Terminal Block
Cable Wire Color
Channel 6 SIG
Yellow
Channel 6 COM
Black/Shield
Channel 6 POWER
Red
Channel 6 ID
Green
Channel 7 SIG
White
Notes:
•
BX-592 is always connected to channel 6 when used for flow control with a APDA-371.
•
BX-592 or BX-596 is required for actual flow control.
•
BX-596 is required for PM
2.5
monitoring, effective March 2007.
Mounting:
The 500 series sensors typically mount near the top of the APDA-371 inlet tube with a short cross-arm and related
hardware. The sensors may also be mounted to a nearby tripod, such as HORIBA model 905.