Operation Manual HORIBA APDA-371
Particulate Monitor
Date:
April, 2010
______________________________________________________________________________________________________________________
______________________________________________________________________________________________________________________
HORIBA Europe GmbH, Julius-Kronenberg-Str. 9, D-42799 Leichlingen,
: +49(0)2175-8978-0,
: +49(0)2175-8978-50
Page 6
1.6
Model APDA-371 PM
2.5
USEPA Equivalent Method
The HORIBA Europe GmbH Model APDA-371 Beta Attenuation Mass Monitor - PM
2.5
FEM Configuration, is designated
as an equivalent method for PM
2.5
monitoring in accordance with 40 CFR Part 53 by the United States Environmental
Protection Agency as of March 12, 2008.
Designation Number: EQPM-0308-170
All of the following parameters and conditions must be observed when the APDA-371 is operated as a PM
2.5
FEM
particulate monitor:
•
The inlet must be equipped with an EPA-designated PM
2.5
Very Sharp Cut Cyclone (VSCC™-A by BGI, Inc.).
The HORIBA stock number for the VSCC™ is BX-808.
•
The inlet must be equipped with a standard EPA PM
10
inlet head. HORIBA BX-802.
•
The unit is operated for hourly average measurements. The PM
2.5
concentration is calculated (external to the
APDA) as a daily average of the hourly concentration measurements made by the APDA-371.
•
The unit must be equipped with firmware revision 3.2.4 or later.
•
The APDA-371 must be operated in proper accordance with this operation manual, revision F or later. A
supplemental BGI Inc. manual is also supplied with the VSCC™.
•
The unit must be equipped with a BX-596 ambient temperature and barometric pressure combination sensor.
This is used for flow control and flow statistics.
•
The unit must be equipped with the internal BX-961 automatic flow controller, and must be operated in Actual
(volumetric) flow control mode.
•
The unit must be equipped with a BX-827 (110V) or BX-830 (230V) Smart Inlet Heater, with the heater RH
regulation set point set to 35%, and Delta-T control disabled.
•
The unit must be equipped with the 8470-1 rev D or later tape control transport assembly with close geometry
beta source configuration. All APDA-371 units manufactured after March 2007 have these features standard.
Older units will have to be factory upgraded and re-calibrated to the latest specifications.
•
The unit must be operated with standard glass fiber filter tape.
•
The COUNT TIME parameter must be set for 8 minutes.
•
The SAMPLE TIME parameter must be set for 42 minutes.
•
The BX-302 zero filter calibration kit is a required accessory. This kit must be used to audit the BKGD
(background) value upon unit deployment and periodically thereafter, as described in the BX-302 manual.
•
The unit may be operated with or without a BX-823 eight foot inlet tube extension and with or without
weatherproof outdoor enclosures BX-902 or BX-903.