pH measurement
26
Tip
You can cancel calibration by pressing the
key while the HOLD icon is blinking.
Note
If calibration of any standard solution is performed again in the calibration mode, only the
value of calibrated solution is updated. If you change to the measurement mode and then
enter the calibration mode to perform calibration again, all previous data is updated.
Confirming the pH electrode status
After calibration is complete, the current pH electrode status is diagnosed from the
calibration result. Use this information for maintenance of the electrode.
Display
Description
Reference
Both
, ERR No.
are not displayed
Electrode sensitivity: 93% to 100%
Good condition.
-
blinks
Electrode sensitivity: 90% to 93%
Urgent measures are not required but attention is
necessary.
lights
Electrode sensitivity: 85% to 90%
Check the electrode immediately.
ERR No.0004
The asymmetry potential is out of the setting range
that allows proper measurement.
Maintain or replace the electrode.
ERR No.0005
The sensitivity is out of the setting range that allows
proper measurement.
Maintain or replace the electrode.