pH measurement
25
5. While the pH electrode is immersed in the standard solution, press the
key.
Stabilization judgment starts and the HOLD icon blinks. When the value is stabilized,
the HOLD icon changes from the blinking state to the lit state and calibration to the set
standard solution value is performed. 1st point calibration ends and the calibration
history icon "1" lights, indicating that the 1st point calibration is complete.
6. Perform the 2nd point calibration. As with the
step 2., wash the pH electrode and immerse it
in the 2nd standard solution. As with the step
4., immerse at least 3 cm from the tip of the pH
electrode.
7. Press the
key and enter the standard solution value for the second point.
8. While the pH electrode is immersed in the standard solution, press the
key.
Stabilization judgment starts and the HOLD icon blinks. When the value is stabilized,
the HOLD icon changes from the blinking state to the lit state and calibration to the set
standard solution value is performed. 2nd point calibration ends and the calibration
history icon "2" lights, indicating that the 2nd point calibration is complete.
Two-point calibration is complete.
When calibrating three points or more, use the same procedure to continue calibrating
the 3rd and subsequent points after the 2nd calibration. You can calibrate up to five
points.
Stabilized
key
Stabilized
key