Honeywell SHK8M Скачать руководство пользователя страница 1

SHK(C)8M(2) : Wireless Shock Sensors -  Installation Instructions (ENGLISH)

Introduction

The SHK8M series is designed to protect window and door

surroundings, and is used to detect forcible attack upon the

surface on which it is mounted.
It is capable to detect 2 different types of forcible attacks : 

Strong attack

:  the device will send an alarm for any single shock

event whose intensity exceed the factory set 

strong attack

sensitivity level.

- Repeated attacks

: the device will send an alarm after a defined

number (controllable) of consecutive shocks, whose intensity

exceed the factory set 

repeated attack

sensitivity level, has been

detected within a period of 8s;
This device is intended for use with a wireless alarm system that

supports the ALPHA or V2GY wireless protocoles such as

Domonial, G2, Dimension, Lynx and Vista control panels.

The SHK8M series

This series consists of the following sensors:
SHK8M: Wireless shock sensor, white.
SHK8M2:  Brown version of the SHK8M.
SHKC8M: Wireless shock sensor featuring a door contact

functionality 
Note: for this sensor the gap between the contact and the magnet

should be less than 10 mm
SHKC8M2: Brown version of the SHKC8M.

Tamper Switches

The SHK8M features two tamper switches to detect device

opening and wrench off from its mounting surface 

Area of Shock Protection

The SHK8M can detect shocks within a radius of 2.5 m. This can

vary, depending on the type of mounting surface. See Figure 1.

Registering the Sensor at the Control Panel

Each sensor  in the SHK8M series has a unique, factory-set serial

number that must be registered at the control panel before it can

be used by the system. Note that the battery must be installed

before registration.
When registering, note the following:
1. Remove the cover of the sensor. See Figure 2.
2. Set SW4 to OFF (Test Mode disabled).
3. Install the battery in the holder provided. See Figure 3.
4. To register, transmit from the unit when prompted, by 

activating its tamper switch as follows: Depress the cover 

tamper switch lightly for several seconds, then open it. The 

panel beeps to confirm registration.
5. Verify the strength of the signal received at the control panel 

(For Domonial, G2 and Dimension only ).

Mounting

1. Remove the cover of the sensor. See Figure 2.
2. Mount the back casing on a solid surface using the screws 

supplied in the three mounting holes. See Figure 4.
For SHKC8M only, mount the magnet (supplied) adjacent to

the mounting plate’s alignment strip. 

See Figure 3.

3. Replace the cover.
Note: Avoid installing the magnet on a metallic surface, such as a

metal door frame.

Sensitivity Configuration

The sensitivity of the device can be adjusted

There are 2 ways to

control the sensitivity:
-  Gross: Select with SW1 between High/lowsensitivity 
-  Fine:  Adjust the sensitivity using the potentiometer. 

In order to verify the correct setting, enter test mode and tap at

the required point within the coverage area shown in figure 1 –

the LED will light up if the threshold is passed.

Note: after any setting change of the DIP switch, device must be

restarted: remove battery, wait 10 seconds, and insert it again to

power on device. 

Pulse Count Configuration

The pulse count is the number of shocks required to trigger an

alarm in case of 

repeated attack

. To configure the pulse count of

the device, you must enter LEARNING MODE. 
1. To enter Learning Mode, power on the sensor with SW 3 in 

the ON position (see Figure 5):
2.   Wait for 10 seconds, until the red LED lights. You will then

have 30 seconds to register the pulse count.   

3. Start tapping the area of coverage – while the red LED is ON.

Each valid impact will turn the LED OFF for 500ms then 

it will return to ON.
The pulse count is automatically determined by the number 

of valid impacts made to the area to be protected  There is no

limit to the number of taps allowed. We recommend to tap 

the area furthest away from the location of the detector.

If the LED doesn’t go off, the area is out of coverage 

range or the impact level is lower than required and the 

shock was not taken into account. Try again using a higher 

sensitivity.
After the 30 seconds the LED will flash to confirm the

number of pulse counts.

4. By default, the pulse count is set to 10. To restore the default 

pulse count, repeat the procedure described in point 1, 

without registering any shocks during the period of 30s after 

the red LED has lit.
At the end of this period, the LED will flash 10 times to 

confirm that the default pulse count has been restored. 

The learn result will be stored into flash memory and will remain

unchanged until the next learning.
Note: In order to calibrate the sensor precisely, you can use the

Viper

calibration tool  device preset to 6 kg/cm

2

constant level of

shock to frame  - never use directly on glass - For areas larger

than 2.5m, it is recommended to use two sensors, in order to

ensure optimal coverage (Ref: PC.06138.00, not supplied).

Test Mode

Use Test Mode to check the configuration and settings of the

sensor. If Learning Mode is disabled or finished and Test Mode

has been enabled (SW 4 in the ON position),  the module enters

Test / Installation Mode. In  Test Mode:
■ The sensor detects the shock, tamper and contact alarm, as 

normal.

■ The LED will indicate an alarm.
■ No RF message will be transmitted
■ Test Mode will be exited after 2 minutes.
Figure 1 on page 3 shows how to use the calibration tool to test

that the sensor is functioning correction. Each shock registered

causes the LED to light. 

Power

The sensor signals a battery failure at the control panel when the

battery needs to be replaced. The sensor will continue to operate

for up to one week after this.  Replace battery with 3V Lithium

battery only.

- 1 -

Содержание SHK8M

Страница 1: ...wsensitivity Fine Adjust the sensitivity using the potentiometer In order to verify the correct setting enter test mode and tap at the required point within the coverage area shown in figure 1 the LED will light up if the threshold is passed Note after any setting change of the DIP switch device must be restarted remove battery wait 10 seconds and insert it again to power on device Pulse Count Con...

Страница 2: ...er le niveau de sensibilité du détecteur Réglage principal sélectionner à l aide de l interrupteur cf figure 5 le niveau de sensibilité souhaité haute basse Réglage fin Pour chaque niveau Haute basse ajuster le niveau de sensibilité avec le potentiomètre cf figure 3 Il est recommandé de régler la sensibilité du détecteur en mode test et de commencer la procédure avec une sensibilité basse Remarque...

Страница 3: ...nstellungsmodi zur Verfügung Grobeinstellung Einstellung der Empfindlichkeit mit SW1 auf hoch oder gering Feineinstellung Einstellung der Empfindlichkeit mit einem Potentiometer Um die Einstellung zu überprüfen geben Sie Testmodus ein und klopfen Sie auf einen bestimmten Punkt innerhalb des Meldebereichs siehe Abb 1 das LED leuchtet auf wenn der Grenzwert überschritten wurde Hinweis Nach jeder Änd...

Страница 4: ...igheid De gevoeligheid van het apparaat kan worden aangepast Er zijn twee manieren om de gevoeligheid te controleren Grof Selecteer met behulp van SW1 tussen hoge lage gevoeligheid Fijn Pas de gevoeligheid aan met behulp van de potentiometer Om de juiste instelling te controleren gaat u in testmodus en tikt u op het vereiste punt binnen het dekkingsbereik dat in Figuur 1 wordt getoond de LED gaat ...

Страница 5: ...ble ajustar la sensibilidad del dispositivo Hay dos formas de controlar la sensibilidad General ponga el SW1 en sensibilidad Alta o Baja Fina ajuste la sensibilidad con el potenciómetro Para confirmar el ajuste correcto acceda al Modo de prueba y toque el punto del área de cobertura que se muestra en la Figura 1 El LED se encenderá cuando se supere el umbral Nota después de los cambios en el inter...

Страница 6: ...ione della sensibilità La sensibilità del sensore può essere regolata La sensibilità può essere regolata in due modi Grossolano Selezionare la sensibilità alta bassa tramite SW1 Fine Regolare la sensibilità utilizzando il potenziometro Per verificare l impostazione attivare il modo prova e urtare un punto compreso nell area protetta indicata nella Figura 1 il LED si accende se la soglia viene supe...

Страница 7: ...on Position de l aimant Magnetposition Locatie van de magneet Ubicación del imán Ubicazione del magnete OFF ON SW1 High sensitivity Sensibilité haute Hohe Empfindlichkeit Hoge gevoeligheid Sensibilidad alta Sensibilità alta low sensitivity Sensibilité basse Geringe Empfindlichkeit Lage gevoeligheid Sensibilidad baja sensibilità bassa SW2 V2 protocol protocole Protokoll Protocollo Alpha protocol pr...

Страница 8: ...l and local regulations In the event of the bat tery being damaged use personal protective equipment to remove it immediately and dispose of it in a safe manner refer to the battery manufacturer s specifications for such situations For Switzerland Annex 4 10 of SR 814 013 applies to the battery included with this product INSTRUCTIONS DE SECURITE CONCERNANT LEREMPLACEMENT ET LAMANIPULATION DES PILE...

Страница 9: ... mm Distanza di chiusura apertura del contatto a lamine 17 21 mm su legno ADVERTENCIA DE SEGURIDAD SOBRE MANIPULACIÓN Y EXTRACCIÓN DE LA BATERÍA Existe un riesgo de incendio explosión y quemaduras la batería no debe recargarse desmontarse calentarse por encima de 100 C ni incinerarse Aunque es seguro extraerla con la mano es preciso ser cuidadosos para no provocar cor tocircuitos entre los bornes ...

Отзывы: