![Honeywell Searchline Excel Скачать руководство пользователя страница 94](http://html.mh-extra.com/html/honeywell/searchline-excel/searchline-excel_technical-handbook_152114094.webp)
94
MAN0530 Issue 10 - 11/09
Searchline Excel
2104M0506
If there are one or more faults present, the display shows:
F-hh:mm DD/MM/YY
eeeeeeeeeeeeeeee
where:
F
Indicates that the message displayed is a fault.
hh:mm
Time the fault occurred.
DD/MM/YY
Date the fault occurred.
eeeeeee...
Fault description. See the following table for a list of error
messages.
See also
Chapter 5, Problem Solving
.
Display Text
Fault Description
Meaning
Bad 4-20mA Loop
ERR_EXCEL_LOOP_FAULT
A fault has occurred in the 4-20mA signal loop.
DSP Fault
ERR_EXCEL_DSP_FAULT
Fault on the gas detector’s digital PCB.
Hardware Fault
ERR_EXCEL_HW_FAULT
A hardware fault has occurred.
NV-RAM Fault
ERR_EXCEL_FRAM_FAULT
A fault has been detected in the unit's NV-RAM.
RTC Fault
ERR_EXCEL_RTC_FAULT
The time and date record stored by the RealTime
Clock (RTC) has been lost.
Software Fault
ERR_EXCEL_SW_FAULT
An gas detector software error has occurred.
Supply Fault
ERR_EXCEL_SUPPLY_FAULT
Incorrect supply voltage reaching the gas
detector.
active Warns
This menu option displays the active warnings in the system. They are displayed in the same
way as described for
Active Faults
except that a
W
is displayed instead of
F
on the first line
of the display.
Display Text
Fault Description
Meaning
Alarm
ERR_EXCEL_LOG_ALARM
The Excel log has become damaged.
Bad Temperature
ERR_EXCEL_TEMPERATURE_LIMIT
Unit has been operated at a temperature
outside its specified
and certified range.
Baseline Drift
ERR EXCEL_DRIFT_LIMIT
Drift of unit’s baseline.
Beam Blocked
ERR_EXCEL_BEAM_BLOCKED
The infrared beam between the Transmitter
and Receiver has become blocked.
Neg Gas Reading
ERR_EXCEL_NEGATIVE_DRIFT
Negative drift of gas detector’s zero position.
Not Installation
ERR_EXCEL_BAD_INSTALLATION
Incorrect gas detector installation has
occurred.
Overrange
ERR_EXCEL_OVERRANGE
An overrange signal has occurred.
Power Failed
ERR EXCEL_LOG_POWER_FAIL
The gas detector has suffered an internal
power failure.
Reset Occurred
ERR_EXCEL_RESET
A gas detector reset has occurred.
Time Adjusted
ERR_EXCEL_LOG_TIME_ADJUST
The Excel log time has been changed.
Uncalibrated
ERR_EXCEL_UNCALIBRATED
The gas detector requires calibration.
Volt Ref Fault
ERR_EXCEL_VOLTAGE_REF_FAIL
Voltage reference fault detected
.
APPENDIX A - HANDHELD INTERROGATOR
Содержание Searchline Excel
Страница 1: ...Searchline Excel Infrared Gas Detectors Technical Handbook ...
Страница 72: ...72 MAN0530 Issue 10 11 09 Searchline Excel 2104M0506 7 CERTIFICATION 7 2 5 System Diagram ...
Страница 79: ...79 MAN0530 Issue 10 11 09 Searchline Excel 2104M0506 7 CERTIFICATION ...
Страница 80: ...80 MAN0530 Issue 10 11 09 Searchline Excel 2104M0506 7 CERTIFICATION With DX100 M Termination Unit ...