Following the procedure described in section
5.1.2
insert a known test filter in the beam
path. Record the response to the test filter and verify that the response is within the
tolerance for stability and repeatability.
•
Following the procedure described in section
5.1.3,
test the unit’s response to high %V/V
gas in a gassing cell. Verify that the response is within the tolerance for stability and
repeatability.