background image

 

29 

 

Click the MODE button once to change the 
selection from hours to minutes (or vice 
versa). 
Hold the MODE button down to advance the 
figure that is currently displayed (hours or 
minutes). 
Once the time has been correctly set, click the 
MODE button three times in rapid succession 
to return to the Main Menu. 

6.4 Set 

date 

The set date submenu is divided into year, 
month and day screens.  The MultiPro will 
automatically scroll through the Year, Month 
and Day screens at 5-second intervals. 

 

 

 

Click the MODE button once to advance 
through the year, month and day screens. 
Hold the MODE button down to advance the 
year, month or day while the specific setting is 
shown. 
Once the date has been correctly entered, 
click the MODE button three times in rapid 
succession to return to the Main Menu. 

6.5 Set 

alarms 

Once the “Set Alarms” 
submenu has been reached, 
the MultiPro will display the 
Default Alarms screen and 
then begin scrolling through 
the various alarm levels. 
Click the MODE button once to advance 
through the screens. 
When the alarm that you want to change is 
shown, hold the MODE button down to 
advance the counter until the appropriate level 
is shown.  If the alarm level required is lower 
than the current setting, allow the counter to 
advance past the maximum possible alarm 
level for the specific channel, and continue 
back to the lower settings. 
To restore the default alarms click the MODE 
button three times in rapid succession with 
“Default Alarms” displayed.  “Def Set” will be 
shown to indicate that the default alarm 
settings have been restored. 
Once the alarm values have been set 
correctly, click the MODE button three times 
rapidly to return to the Main Menu. 

Note:  To enable the H

2

S TWA alarm setting 

of 1.6 PPM, first enable the H

2

S decimal point 

as described above in section 6.2.  The 1.6 
PPM alarm value will then appear as an 
option in the TWA alarm setting for H

2

S. 

6.6 

Set cal gas 

Once the set cal gas submenu has been 
reached, the MultiPro will scroll through the 
various calibration gas values screens and the 
set default cal gas option. 
Click the MODE button once to advance 
through the screens. 
Hold the MODE button down to advance the 
counter for a specific calibration gas. 
To set default calibration gas 
levels click the MODE button 
three times in rapid 
succession with “Default Cal 
Gas” displayed. 
Once the calibration gas values have been set 
correctly, click the MODE button three times in 
rapid succession to return to the main menu. 

6.7 

Set cal due 

Once the set cal due 
submenu is reached, the cal 
due screen will be shown. 
Hold the MODE button down 
to advance the counter.  The number of days 
shown is the number of days that must pass 
following a successful calibration before the 
calibration due warning will be shown.  The 
maximum length of time between calibrations 
is 180 days. 
To disable the calibration due warning, set the 
interval to 0 days (OFF). 
Once the calibration due interval has been set 
correctly, click the MODE button three times 
rapidly to return to the Main Menu. 

Note:  Honeywell Analytics recommends 
regular verification of accuracy with 
calibration gas.  See Appendix B for 
details. 

6.8 

Set bump test due 

Once the bump due submenu is reached, the 
bump due screen will be shown.  The number 
of days shown is the number of days that 
must pass following a successful calibration 
before the bump due warning will be shown.  
Hold the MODE button down to advance the 
counter in days.  OFF will be followed by 
DAILY, which will then be followed by 2 days 
and so on.  Once the interval passes 7 days, it 

Содержание MultiPro

Страница 1: ......

Страница 2: ... IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED MultiPro Reference Manual Part Number 13 278 Version 3 Copyright 2012 by Honeywell Analytics Inc Lincolnshire Illinois 60069 All rights reserved No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above Honeywell Analytics res...

Страница 3: ...n using the MultiPro pump 11 1 8 Data storage 11 1 8 1 Black box data recorder 11 1 8 2 Event logger 11 1 9 MultiPro design components 11 1 10 MultiPro standard accessories 12 1 10 1 Alkaline MultiPro detectors 12 1 10 2 Li Ion MultiPro detectors 12 1 11 MultiPro kits 12 1 11 1 MultiPro Confined Space Kits 12 1 11 2 MultiPro Value Packs 12 2 BASIC OPERATIONS 12 2 1Turning the MultiPro On 12 2 1 1 ...

Страница 4: ...lines for the Li Ion versions 24 5 5 2 Charging guidelines for Li Ion battery 25 5 5 3Charging procedure for Li Ion battery 25 5 5 4 Charging with the pump attached 25 5 5 5 Battery troubleshooting 25 5 6 Sensor replacement 25 5 6 1 Sensor replacement Not Duo Tox 25 5 7 Sample probe assembly 26 5 7 1 Changing sample probe filters 27 5 7 2 Changing sample probe tubes wands 27 5 8 MultiPro Pump Main...

Страница 5: ...Appendix D Toxic Sensor Cross Sensitivity 33 Appendix E Basic Parts List 33 MultiPro Remote Sampling Accessories 33 MultiPro Sensors 33 MultiPro Charging Datalogging Accessories 33 Appendix F Honeywell Analytics Standard Gas Detection Warranty 35 ...

Страница 6: ...entially hazardous situation which if not avoided could result in death or serious injury indicates a potentially hazardous situation which if not avoided may result in moderate or minor injury CAUTION used without the safety alert symbol indicates a potentially hazardous situation which if not avoided may result in property damage Warnings and Cautions 1 The MultiPro personal portable gas detecto...

Страница 7: ...ere that is known to contain 20 9 oxygen 0 0 LEL and 0 PPM toxic gas 11 The accuracy of the MultiPro should be checked immediately following any known exposure to contaminants by testing with known concentration test gas before further use Failure to check accuracy can lead to inaccurate and potentially dangerous readings 12 The accuracy of the MultiPro should be checked immediately following any ...

Страница 8: ...eference manual before operating or servicing the MultiPro 18 A rapid up scale reading followed by a declining or erratic reading may indicate a hazardous combustible gas concentration that exceeds the MultiPro s zero to 100 percent LEL detection range 19 The MultiPro is not designed for use in oxygen enriched atmospheres 20 Do not use the MultiPro pump for prolonged periods in an atmosphere conta...

Страница 9: ... the end of the sample draw probe Use of the hand aspirated sample draw kits is covered in section 3 1 Use of the motorized sample draw pump is covered in section 3 2 A detailed description of the MultiPro probe assembly is given in section 5 7 1 2 Multi sensor capability The MultiPro can be configured to simultaneously monitor oxygen carbon monoxide hydrogen sulfide and combustible gases and vapo...

Страница 10: ... to immediately leave the affected area and to return only after further testing determines that the area is once again safe for entry Failure to immediately leave the area may result in serious injury or death A rapid up scale reading followed by a declining or erratic reading may indicate a hazardous combustible gas concentration that exceeds the MultiPro s zero to 100 percent LEL detection rang...

Страница 11: ...ously for proper voltage The MultiPro also monitors the connection of sensors that are currently installed The detection of any electronic faults causes the activation of the audible and visible alarms and causes the display of the appropriate explanatory message 1 5 1 Heartbeat At the center of the display is a heart symbol that will blink every few seconds to show that the instrument is function...

Страница 12: ...ro will begin to write the new data over the oldest data The black box data recorder will store approximately 41 hours of data in one minute increments before it begins to write new data over the oldest data In this way the newest data is always conserved To extract the information from the black box data recorder the MultiPro must be returned to Honeywell Analytics Once the data is downloaded fro...

Страница 13: ... ten feet of tubing The sample probe itself is also available separately Standard configurations of the MultiPro are delivered in a cardboard box with cardboard inserts 1 10 1 Alkaline MultiPro detectors If the MultiPro has been purchased as an alkaline instrument the standard accessories include a set of 3 disposable AA alkaline batteries 1 10 2 Li Ion MultiPro detectors If the MultiPro has been ...

Страница 14: ...Pro instruments equipped with a pump see section 2 1 1 below For instructions on changing the time and date with the MODE button see section 6 below The instrument temperature will then be shown The warning alarm levels screen will then be shown followed by the danger STEL and TWA alarm levels screens For more information concerning STEL and TWA alarms see sections 2 4 3 and 2 4 4 After the alarm ...

Страница 15: ...D once every 5 seconds as a reminder If the MultiPro is due for span calibration the calibration bottle icon will be shown Both 0 Cal and the calibration bottle icon will be shown if the instrument is due for both span and fresh air calibration If a sensor is not detected in one of the sensor channels during start up the reading in the designated sensor channel will appear blank If a complete sens...

Страница 16: ...ence 2 4 3 STEL Alarms STEL Short Term Exposure Limit alarm values represent the average concentration of instrument readings for the target gas for the most recently completed 15 minutes of operation The default STEL alarm level for the MultiPro CO sensor is 100PPM The default STEL alarm value for the MultiPro H2S sensor is 15PPM 2 4 4 TWA Alarms TWA Time Weighted Average values are calculated by...

Страница 17: ...ontaining 20 9 oxygen 0 LEL and 0 PPM toxic gases and then turned back on The display will show and OL alternately in place of the sensor reading for any channel that has gone into over range alarm A combustible sensor overrange alarm indicates a potentially explosive atmosphere Failure to leave the area immediately may result in serious injury or death In the event of an LEL overrange alarm the M...

Страница 18: ...he bottom of the instrument The sensor ports must be kept free of obstruction Blocked sensor ports can lead to inaccurate and potentially dangerous readings In diffusion mode the atmosphere being measured reaches the sensors by diffusing through vents in the bottom of the instrument Normal air movements are enough to carry the sample to the sensors The sensors react quickly to changes in the conce...

Страница 19: ...ds free remote monitoring Use of the motorized sample draw pump allows the MultiPro to continuously monitor remote locations The pump is powered by the MultiPro battery When the pump is attached to the instrument PUMP will be shown on the display in the current gas readings screen Note The maximum amount of tubing that can be used with the motorized sample draw pump is 50 feet To ensure accurate r...

Страница 20: ...ump is designed to detect pressure changes while the sample draw probe is being held in a vertical position If the probe is held horizontally or at a low angle while inserted into a fluid a pressure drop sufficient to cause the pump to shut down may not be generated and water could be drawn into the pump assembly causing damage to the pump sensors and internal components of the MultiPro CAUTION In...

Страница 21: ...or further details on sensor stabilization requirements 2 Make sure the instrument is located in fresh air Figure 4 1 Bump Test Span Calibration set up 3 Verify that the current gas readings match the concentrations present in fresh air The oxygen O2 sensor should read 20 9 0 2 vol The readings for the LEL sensor should be 0 LEL and toxic sensors should read 0 or 0 0 parts per million PPM in fresh...

Страница 22: ...en until a successful fresh air calibration is performed If a specific sensor has failed the fresh air zero calibration it will be displayed with the 0 Cal icon If a successful fresh air zero calibration is not performed prior to instrument shut down the MultiPro will show that Fresh Air Calibration is due during instrument start up Possible causes and solutions 1 The atmosphere in which the instr...

Страница 23: ...lytics calibration gas cylinders typically contain approximately 18 0 oxygen In order to pass the span calibration the instrument must register an oxygen reading below 19 5 during span calibration See section 4 3 2 below if the oxygen sensor does not detect the drop in oxygen level and fails the span calibration The calibration is fully automatic from this point on Upon successful calibration of a...

Страница 24: ...anner as the toxic and LEL sensors If the O2 sensor fails to register a reading below 19 5 during the span calibration the display will show O2 Too Low followed by the O2 CAL Error screen immediately after the failed calibration attempt Press MODE to acknowledge the warning and turn the instrument off If the oxygen sensor fails to register the drop in oxygen during the span calibration while being...

Страница 25: ...dous area may impair intrinsic safety Use only Duracell MN1500 or Ultra MX1500 Eveready Energizer E91 LR6 Eveready EN91 batteries Substitution of batteries may impair intrinsic safety To replace the alkaline batteries 1 Remove the battery pack from the MultiPro as discussed in above in section 5 3 2 Remove the three alkaline batteries and replace them Be sure to align the positive and negative end...

Страница 26: ...e charger and with the guide pins facing down towards the charger as shown in the illustration at right Once the spacer has been inserted the MultiPro with pump may be charged according to the instruction given in section 5 5 3 above 5 5 5 Battery troubleshooting If the green Charge LED on the charger fails to light when the MultiPro with Li Ion battery pack is placed in the charger remove the ins...

Страница 27: ...d 10 assume that the sensor stabilization period has passed 9 If an oxygen sensor was replaced perform the Fresh Air Zero calibration as discussed in section 4 2 10 If a combustible LEL or a toxic sensor has been replaced perform both the Fresh Air Zero calibration and the Span calibration as discussed in sections 4 2 and 4 3 5 7 Sample probe assembly The MultiPro s sample draw probe is the standa...

Страница 28: ...e tubes are exchanged or replaced before being returned to service 5 8 MultiPro Pump Maintenance MultiPro pumps are fairly maintenance free with the exception of the replacement of the pump filters on a regular basis 5 8 1 Replacing pump filters 1 Remove the two screws that hold the inlet port to the pump 2 Gently pull the dust filter holder free of the pump 3 Remove and replace the dust filter th...

Страница 29: ...ubmenu is entered the MultiPro will automatically scroll through the options at 5 second intervals To change a setting for on to off or vice versa hold the MODE button down for 1 2 seconds Once the settings have been modified as needed click the MODE button three times in rapid succession to return to the Main Menu STEL and TWA Alarms are discussed in Appendix A The Security Beep is covered in sec...

Страница 30: ...nu Note To enable the H2S TWA alarm setting of 1 6 PPM first enable the H2S decimal point as described above in section 6 2 The 1 6 PPM alarm value will then appear as an option in the TWA alarm setting for H2S 6 6 Set cal gas Once the set cal gas submenu has been reached the MultiPro will scroll through the various calibration gas values screens and the set default cal gas option Click the MODE b...

Страница 31: ...val and Bump Test Reminder are used exclusively with the MultiPro IQ Express Dock 6 9 Setup accept Once the settings in any of the submenus have been changed as needed return to the Main Menu by clicking the MODE button three times in rapid succession Once in the Main Menu hold the MODE button down for three seconds to exit The MultiPro will provide a countdown to accept changes to the settings Cl...

Страница 32: ...ects due to chronic exposure at lower levels that take the greatest toll of affected workers This is the situation with regards to smokers parking garage attendants or others chronically exposed to carbon monoxide in the workplace Exposure levels are too low to produce immediate symptoms but small repeated doses reduce the oxygen carrying capacity of the blood over time to dangerously low levels T...

Страница 33: ...uidelines concerning calibration frequency Honeywell Analytics procedures closely follow these guidelines If your operating procedures do not permit daily checking of the sensors Honeywell Analytics recommends the following procedure to establish a safe and prudent accuracy check schedule for your Honeywell Analytics instruments 1 During a period of initial use of at least 10 days in the intended ...

Страница 34: ...NH3 C2H4 C2H2 Carbon Monoxide CO 100 10 5 10 15 5 15 50 15 3 0 75 250 Hydrogen Sulfide H2S 0 5 100 20 2 20 20 60 0 2 0 0 0 n d n d Appendix E Basic Parts List MultiPro Remote Sampling Accessories Part No Description 54 49 102 Continuous sample pump for MultiPro Includes 10 of sample tubing and sample probe assembly UL CSA approved version 54 49 102 5 Continuous sample pump for MultiPro Includes 10...

Страница 35: ...ropean 54 49 107 MultiPro Li Ion rechargeable battery pack UL CSA Approved version 54 49 107 5 MultiPro Li Ion rechargeable battery pack ATEX approved version European 54 26 0603 MultiPro BioTrak Software Kit Includes BioTrak software and manual 54 26 0605 Infrared communication device Serial IrDA Requires one available PC serial port ...

Страница 36: ...FITNESS FOR A PARTICULAR PURPOSE HONEYWELL ANALYTICS WILL NOT BE LIABLE FOR LOSS OR DAMAGE OF ANY KIND CONNECTED TO THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY Instrument Accessory Warranty Periods Product s Warranty Period PHD6 2 years from date of purchase ToxiPro MultiPro 2 years from date of purchase Battery packs and chargers sampling pumps and other com...

Отзывы: