Use only Duracell MN1500 or Ultra MX1500, Eveready Energizer
E91-LR6, Eveready EN91 batteries in the alkaline battery pack. Substitution of
batteries may impair intrinsic safety.
5
.
To reduce the risk of explosion, do not mix old or used
batteries with new batteries and do not mix batteries from different manufacturers.
6.
Do not charge the MultiPro with any charger other than the
appropriate MultiPro charger. Standard versions of the MultiPro must be charged
with the UL/CSA-approved charger, which is part number 54-49-103-1. European
versions of the MultiPro must be charged with the ATEX-approved charger, which is
Honeywell Analytics part number 54-49-103-5.
7.
The MultiPro must be located in a non-hazardous location
during the charging cycle. Charging the MultiPro in a hazardous location may
impair intrinsic safety.
8.
MultiPro rechargeable battery packs are supplied with
Panasonic CGA103450 Lithium-Ion batteries. These batteries are not replaceable by
the user. The rechargeable pack must be obtained from Honeywell Analytics and
replaced as an assembly.
9.
The accuracy of the MultiPro should be checked periodically
with known concentration calibration gas. Failure to check accuracy can lead to
inaccurate and potentially dangerous readings. (The Canadian Standards
Association (CSA) requires an accuracy check using known concentration
calibration gas prior to each day’s use.)
10.
Fresh air/zero calibrations may only be performed in an
atmosphere that is known to contain 20.9% oxygen, 0.0% LEL and 0 PPM toxic gas.
11.
The accuracy of the MultiPro should be checked immediately
following any known exposure to contaminants by testing with known concentration
test gas before further use. Failure to check accuracy can lead to inaccurate and
potentially dangerous readings.
12.
The accuracy of the MultiPro should be checked immediately
following any physical shock. Failure to check accuracy can lead to inaccurate and
potentially dangerous readings.
13.
A sensor that cannot be calibrated or is found to be out of
tolerance should be replaced immediately. An instrument that fails calibration may
not be used until testing with known concentration test gas determines that
accuracy has been restored, and the instrument is once again fit for use.
14.
Do not reset the calibration gas concentration unless you are
using a calibration gas concentration that differs from the one that is normally
supplied by Honeywell Analytics for use in calibrating the MultiPro.
Customers are strongly urged to use only calibration materials when calibrating the
MultiPro. Use of non-standard calibration gas and/or calibration kit components can
lead to dangerously inaccurate readings and may void the standard Honeywell
Analytics warranty.
15.
Use of non-standard calibration gas and/or calibration kit
components when calibrating the MultiPro can lead to inaccurate and potentially
dangerous readings and may void the standard Honeywell Analytics warranty.
Honeywell Analytics offers calibration kits and long-lasting cylinders of test gas
specifically developed for easy MultiPro calibration. Customers are strongly