GENESIS
TM
DM MONITOR
Programming/Troubleshooting Guide
P4520EN.04 07/14/2010
4
After transmitting vitals, pressing
MANUAL
twice clears vitals and returns the display to idle mode.
If
MANUAL
is not pressed the vitals will be presented for 5 minutes and then cleared automatically.
Setup Mode
Quickly press and release the
START
and
YES
keys simultaneously.
DO NOT hold the keys down.
Setup Mode normally opens to the
Time
Menu
. However, if you press and hold instead of press and
release, you may scroll to a different menu.
To press a key correctly, press firmly in the middle of the key. When you press correctly, you
will hear a beep.
If you are in Setup Mode during a scheduled report time, the monitor will skip the monitoring
session and will not transmit any information. LifeStream will post a Transmit Error (orange alert) for
the missed reading.
Scheduling options
OFF
Every day (Su Mo Tu We Th Fr Sa)
Any combination of days ( _ _ Mo _ _ We _ _ Fr _ _)
Every time (Su)
Only once per day (S*)
If you choose one time per day (day marked with an asterisk:
M
*), the monitor collects the
scheduled data during the first monitoring session whether initiated by a report time or by
pressing
RETEST
. Once the scheduled data is collected, the monitor will not collect it again until
the next day. The monitor resets to a new day at midnight.
If the monitor times out without a patient response, it will attempt to collect the scheduled data
the next monitoring session. Timing out or powering down will not affect the once-a-day
schedule.
Downloading a monitor configuration
After an initial configuration packet has been sent from the monitor, a nurse or other authorized person
can remotely download a configuration from LifeStream to the monitor.
To download a configuration from a monitor, press
MANUAL
. When the manual mode acquisition is
displayed, press YES. This will start a configuration download through the phone line.