GENESIS

TM

 DM MONITOR

 

Programming/Troubleshooting Guide 

 

P4520EN.04 07/14/2010

   

 

 

 

 

25

 

VFD Display 

Meaning 

Possible Cause 

Resolution/Action 

617 

MODEM NOT READY 

Modem was not ready. 

 

If problem persists, call 
Honeywell HomMed. 

901 

RAM_ERROR 

RAM self test failed 

 

Call Honeywell HomMed. 

902 

ROM_ERROR 

ROM Checksum error 

 

Call Honeywell HomMed. 

903 

UART_ERROR 

Hardware error 

 

Call Honeywell HomMed. 

904 

RTCC_ERROR 

Real time clock failure 

 

Call Honeywell HomMed. 

910 ERROR_FILE_SYSTEM 

Language text files missing 
or corrupt 

 

Replace the SmartMedia card. 

911 

DISK_ERROR 

SmartMedia failure 

 

Replace the SmartMedia card. 

912 DISK_ERROR_BAD_FORMAT  SmartMedia failure – bad 

format 

 

Replace the SmartMedia card. 

913 DISK_ERROR_NO_CARD 

SmartMedia failure – card 
removed 

 

Insert the correct SmartMedia 
card. 

914 DISK_ERROR_BAD_CARD 

SmartMedia failure – bad 

card 

 

Replace the SmartMedia card. 

915 DISK_ERROR_CHANGED_CARD SmartMedia failure – card 

swapped 

 

Reboot the unit; if problem 
persists, call Honeywell HomMed. 

916 DISK_ERROR_CARD_FAILURE  SmartMedia failure – card 

failure 

 

Replace SmartMedia card; if 

problem persists, call Honeywell 
HomMed. 

930 PHOTOCARD 

TIMEOUT 

No response – card 
removed or defective 

Card removed before monitor 
could download images. 

Defective card 

Retrain the patient. 
 

 

Use new card, retake photo, 

resend. 

931 PHOTOCARD 

FAILURE 

Card not readable: 
damaged, wrong type 

 

Use only SmartMedia cards for 
digital images. 

Use new card, retake photo, 

resend. 

932 PHOTOCARD 

READ 

ERROR 

Invalid CRC, parity error, 
etc. 

 

Take another picture and resend. 

Use new card, retake photo, 

resend. 

933 

PHOTOCARD CONFIG BAD 

Config. file (*.CFG) missing 
or contains invalid info 

 

Use only appropriate SmartMedia 
cards for digital images. 

Use new card, retake photo, and 

resend. 

934 

PHOTOCARD PAT. ID BAD 

Pt. ID file (*.PAT) missing or 
contains invalid info 

 

Use new card, retake photo, and 
resend. 

935 

PHOTO NOT FOUND 

No photos in photo 

directory 

 

Take pictures and resend. 

 
 

Содержание HomMed Genesis

Страница 1: ...oneywell HomMed Customer Service Toll free 888 353 5404 Phone 262 252 5840 Fax 262 252 6105 Honeywell HomMed Corporate Offices Honeywell HomMed 3400 Intertech Drive Suite 200 Brookfield WI 53045 Phone 262 783 5440 Fax 262 252 6105 Copyright 2010 Honeywell HomMed All rights reserved ...

Страница 2: ...mer Service will make every attempt to resolve the issue over the phone Special Conventions The following table lists the conventions used in this guide This convention Identifies Function font Keys that you press on the keyboard sign between key names Keys that you press simultaneously START YES Courier font Text that displays in the monitor display Italics Menu names publication titles or refere...

Страница 3: ... 4 Gray MANUAL key Gray MANUAL key enters Manual Mode Exits a menu in Setup Mode 5 Weight Display Displays weight from attached scale in pounds or kilograms as configured 6 Blue RETEST key Blue RETEST key starts an unscheduled monitoring session Selects a menu option in Setup Mode 7 Black NO key Black NO key records a no answer DOWN scroll key in Setup Mode 8 White YES key White YES key records a ...

Страница 4: ...duling options OFF Every day Su Mo Tu We Th Fr Sa Any combination of days _ _ Mo _ _ We _ _ Fr _ _ Every time Su Only once per day S If you choose one time per day day marked with an asterisk M the monitor collects the scheduled data during the first monitoring session whether initiated by a report time or by pressing RETEST Once the scheduled data is collected the monitor will not collect it agai...

Страница 5: ...GENESISTM DM MONITOR Programming Troubleshooting Guide P4520EN 04 07 14 2010 5 Programming menu chart The programming menu chart below illustrates the location of each of the monitor menus ...

Страница 6: ...ing your inhalers more than usual Increased need for inhalers signals n a change in condition Use inhalers AS PRESCRIBED If difficult to breathe chest tight or in pain or sudden cough that continues Contact your CARE PROVIDER 04 Did you need extra pillows to sleep comfortably last night Using extra pillows to sleep may n signal change Use oxygen inhalers AS PRESCRIBED Contact your CARE PROVIDER EX...

Страница 7: ...aving any headaches 15 Have your ankles been swollen more than usual Ankle swelling caused by What you eat medicines you take Elevate legs Limit salt IF ORDERED Call CARE PROVIDER Feet or legs more swollen 16 Are you having any difficulty standing or walking Difficulty standing or walking At risk for falls Learn cause what to do Discuss with Care Provider 17 Have you been having any new numbness o...

Страница 8: ...g difficulty taking care of your wound 33 Have your hands or face been more swollen in the last day 34 Has there been any protein in your urine in the last day 35 Do you have severe heartburn today 36 Have you noticed a decrease in your baby s movement today 37 Have you had any blurred vision today 38 Have there been any changes in the medication you are taking 39 Are you having difficulty followi...

Страница 9: ...ing your activities may signal a change in your medical condition If you are unable to resume normal activities have difficulty breathing or suddenly develop a new cough contact your health care provider 54 Are you having difficulty taking any of your medications same as Question Number 8 Difficulty taking medications taking medications at irregular times or not taking medications consistently may...

Страница 10: ...uld be discussed with your health care provider 60 Have you developed a cough same as Question Number 28 A newly developed cough may signal a change in your medical condition If your cough is unrelieved with prescribed medication or if you develop difficulty breathing contact your health care provider 61 Have you had a fall in the last day same as Question Number 31 Falls can be a major cause of i...

Страница 11: ...a meeting this week 112 Remember your grandchildren are coming over this week have fun 113 Please remember to change your dressing today 114 Please remember to keep feet and legs elevated while sitting 115 Please remember to eat foods low in Sodium 116 Please remember to wear your oxygen as prescribed 117 Please remember to use your nebulizer as prescribed 118 Please remember to do your exercises ...

Страница 12: ...early every day during your sleep 7 Has anyone noticed that you quit breathing during sleep 3 4 times a week 8 Do you feel tired or fatigued today after your sleep 9 Do you feel tired fatigued or not up to par during your wake time nearly every day 10 Do you feel tired fatigued or not up to par during your wake time 3 4 times a week 11 Do you nod off or fall asleep while driving a car nearly every...

Страница 13: ... the wall jack 2 Insert duplex jack 3 Plug caller ID box phone cord into duplex jack 4 Plug the monitor phone cord into duplex jack Patient already has a duplex jack in the jack outlet Replace with triplex jack 1 Remove duplex jack and insert triplex jack 2 Plug patient s equipment into two of the jack outlets and use third outlet for the monitor Have two short phone cords but need a longer cord U...

Страница 14: ...urface 4 Reattach the scale cable to the scale 5 Reboot the monitor Scale does not zero properly monitor displays dashes instead of 0 0 If scale still does not zero it must be recalibrated Call Honeywell HomMed Customer Service Patient decompensates easily after obtaining weight The oximeter obtains an SpO2 reading in either 8 beats or 8 seconds The reading is taken again if it is out of range on ...

Страница 15: ...rrectly positioned Poor patient perfusion Fingernail polish or false nails are interfering with sensor light Defective sensor or cable 1 Plug oximeter into plug receptacle and make sure it firmly latches into place 2 Instruct the patient on proper placement 3 Reposition sensor on patient 4 Have patient remove finger nail polish or false nails 5 Replace finger sensor Heart rate is very low or does ...

Страница 16: ...gramming the monitor Verify that the cable is firmly connected to the monitor as well as the device No data transferred from device to monitor Monitor does not have a port programmed for the device Assign a com port to the device using the Option menu Communication errors between device and monitor Noise Cable problems Place the monitor so that it is not on near an electrical device such as a TV o...

Страница 17: ... voice communications over a network such as the Internet or other packet switched networks Other terms frequently encountered and synonymous with VOIP include but are not limited to IP telephony Internet telephony voice over broadband VoBB broadband telephony and broadband phone Internet telephony refers to communications services voice facsimile and or voice messaging applications that are trans...

Страница 18: ... numbers Verify that Transmit Mode is programmed correctly for your area If you do not know the correct mode contact Customer Service Monitor collected the vitals but did not transmit Wall jack does not work Faulty duplex jack Monitor serial number does not match unit label Monitor must dial a prefix for an outside line Facility has digital line Test wall jack with a phone Use a new duplex jack Go...

Страница 19: ... SCALE_NOT_ CONFIGURED Scale not configured or configuration not sensed Scale not configured through the Setup Wizard but is turned ON in the Menu Settings Re boot re set the system first power OFF both the Scale and the DM Monitor next power ON the DM monitor first then the Scale 210 NIBP_TIMEOUT User timeout NIBP Patient noncompliance Retrain for patient 211 NIBP_COMM Lost communication with NIB...

Страница 20: ...rror Call Honeywell HomMed 244 TEMP_RECOVERABLE Temperature error Probe not positioned properly Retrain the patient 245 TEMP_NO_PROBE Temperature probe disconnected Connect the probe Verify all connections secure 246 TEMP_HI_AMBIENT Temperature probe too warm Operating range 84 to 108 F 28 9 to 42 2 C 247 TEMP_LO_AMBIENT Temperature probe too cold Operating range 84 to 108 F 28 9 to 42 2 C 250 IDC...

Страница 21: ...ove the block to allow the monitor to send data by modem 305 MODEM_BAUD_RATE Baud rate lock failed If using remote gateway transmission method ensure the number is correct and the remote gateway host is properly configured If problem persists call Honeywell HomMed 311 Modem Logon Skytel logon not accepted If problem persists call Honeywell HomMed 312 Modem Disconnect Skytel forced disconnect If pr...

Страница 22: ...e 412 GLUCO_OFF Glucose meter not enabled In Options menu set appropriate glucose meter to On 413 WRONG_GLUCO ID Card and monitor setup gluco type mismatch Program the card or the monitor for the correct glucose meter 420 SPIRO_TIMEOUT User timeout spirometer Patient noncompliance Retrain the patient 421 SPIRO_COMM Spirometer disconnected Spirometer cable not connected Loose connection Connect the...

Страница 23: ...ists call Honeywell HomMed 603 Call Customer Service User name or password was incorrect The monitor has not been activated with the ISP by Honeywell HomMed staff The ISP configuration has been modified and is no longer correct Internal and external serial number do not match If problem persists call Honeywell HomMed 604 Call Customer Service Failed communications with modem If problem persists ca...

Страница 24: ... properly DSL is not configured properly The ISP is not properly accepting calls Verify the ISP1 Phone and ISP2 Phone number are configured properly in the ISP Settings menu If problem persists call Honeywell HomMed 612 ISP Login Failed Failed to log into the ISP The phone number is not configured properly The ISP is not properly accepting calls Verify the ISP1 Phone and ISP2 Phone number are conf...

Страница 25: ...dia card 915 DISK_ERROR_CHANGED_CARD SmartMedia failure card swapped Reboot the unit if problem persists call Honeywell HomMed 916 DISK_ERROR_CARD_FAILURE SmartMedia failure card failure Replace SmartMedia card if problem persists call Honeywell HomMed 930 PHOTOCARD TIMEOUT No response card removed or defective Card removed before monitor could download images Defective card Retrain the patient Us...

Страница 26: ...r Resources Genesis DM User Guide Honeywell HomMed LLC 2009 Genesis DM Monitor Quick Setup Guide Honeywell HomMed 2007 LifeStream User Guide Honeywell HomMed 2009 For more information about Honeywell HomMed authorized peripheral devices refer to the appropriate manufacturer s material ...

Отзывы: