VFD Display
Meaning
Possible Cause
Resolution/Action
451
PEAKFLOW_COMM
Peak flow discontinued
Peak flow cable not connected
Loose connection
Connect the cable.
Make sure all connections are
secure
452 PEAKFLOW_OFF
Peak flow not assigned a
COM port
In Options menu, set peak flow
to On.
601
Relay URL Rejected
Remote TCP Server refused
connection
The HomMed Relay is having a
problem and refused to accept a
connection to the monitor.
The monitor is not properly
configured for what HomMed
relay servers to use (not a
clinician accessible configuration
item)
If problem persists, call
Honeywell HomMed.
If problem persists, call
Honeywell HomMed.
602 Relay
URL
Rejected
Connection failed for
unknown reason
The HomMed Relay is having a
problem and fails to respond to
monitor in a timely manner.
The monitor is not properly
configured for what HomMed
relay servers to use (not a
clinician accessible configuration
item)
If problem persists, call
Honeywell HomMed.
If problem persists, call
Honeywell HomMed.
603
Call Customer Service
User name or password
was incorrect
The monitor has not been
activated with the ISP by
Honeywell HomMed staff.
The ISP configuration has been
modified and is no longer correct.
Internal and external serial
number do not match.
If problem persists, call
Honeywell HomMed.
604
Call Customer Service
Failed communications with
modem
If problem persists, call
Honeywell HomMed.
605 Call
Interrupted
Communications with
server failed after
connection established
Call waiting is active and an
incoming call has interrupted the
transmission.
If problem persists, call
Honeywell HomMed.
606 CONNECTION
CLOSED
Socket was closed by
remote host
If problem persists, call
Honeywell HomMed.
607
PIN NUMBER DENIED
Pin number not recognized
by server
The pin number is not correctly
configured on the monitor.
The pin number is not yet
activated on the HomMed relay
server.
Verify the pin number is correct.
If problem persists, call
Honeywell HomMed.
608
ISP BUSY SIGNAL
Modem received a busy
signal
The phone number is not
configured properly.
An 800/900 block is enabled
The ISP is not properly accepting
calls.
Verify the ISP1 Phone # and ISP2
Phone number are configured
properly.
If problem persists, call
Honeywell HomMed.