GENESIS
TM
DM MONITOR
Programming/Troubleshooting Guide
P4520EN.04 07/14/2010
22
#
VFD Display
Meaning
Possible Cause
Resolution/Action
324 PAGER_MSG_FAIL
Message
failed
If problem persists, call
Honeywell HomMed.
350 GPRS
Requires
Reset
Modem responded with an error
the monitor doesn't handle, too
early to be more specific with
causes here
If problem persists, call
Honeywell HomMed
351
Call Customer Service
Improper Hardware Setup
SIM card is missing or not
enabled
If problem persists, call
Honeywell HomMed
352
GPRS Service Unavailable
Signal is too weak for the
modem to work
No service in the area
If problem persists, call
Honeywell HomMed.
353
GPRS Network Fail
Access Point Name is not
configured correctly on the
modem
No service in the area
If problem persists, call
Honeywell HomMed.
410 GLUCO_TIMEOUT
User timeout – glucose
meter
Patient noncompliance
Retrain the patient.
411
GLUCO_COMM
Glucose meter disconnected
Glucose meter cable not
connected
Loose Connection
Connect the cable.
Make sure all connections are
secure.
412
GLUCO_OFF
Glucose meter not enabled
In Options menu, set appropriate
glucose meter to On.
413 WRONG_GLUCO
ID Card and monitor setup
gluco-type mismatch
Program the card or the monitor
for the correct glucose meter.
420
SPIRO_TIMEOUT
User timeout - spirometer
Patient noncompliance
Retrain the patient.
421 SPIRO_COMM
Spirometer
disconnected
Spirometer cable not connected
Loose connection
Connect the cable.
Make sure all connections are
secure.
422 SPIRO_OFF
Spirometer not enabled (no
assigned COM port)
In Options menu, set appropriate
spirometer to On.
430
PT_TIMEOUT
User timeout – PT/INR
Patient noncompliance
Retrain the patient.
431 PT_COMM
PT/INR
disconnected
PT/INR cable not connected
Loose connection
Connect the cable.
Make sure all connections are
secure.
432 PT_OFF
PT/INR not enabled (no
assigned COM port)
In Options menu, set appropriate
PT/INR to On.
440
ERROR_ECG_TIMEOUT
User timeout – ECG
Patient noncompliance
Retrain patient.
441 ERROR_ECG_COMM
ECG
disconnected
ECG cable not connected
Loose connection
Connect cable.
Make sure all connections are
secure.
442 ERROR_ECG_OFF
ECG not assigned a COM
port
In Options menu, set ECG to On.
450
PEAKFLOW_TIMEOUT
User timeout – peak flow
Patient noncompliance
Retrain the patient.