GasAlert Extreme
User Manual
39
Setting the High Alarm Setpoint
Note
To bypass and retain the current high alarm setpoint value,
press
C
. The detector then returns to the normal operation.
When the low alarm setpoint value has been changed or bypassed, the
Set HIGH
alarm setpoint screen displays.
18. Press
D
or
E
to scroll to the required value.
19. Press
C
to save the new value and return to normal operation.
Note
If a new value is selected but not confirmed within 10 seconds
by pressing
C
, the detector automatically retains the original
value and
NO
displays. The detector proceeds to normal oper-
ation.
When calibration is complete, the detector beeps and vibrates four
times, and the LEDs flash four times before returning to normal opera-
tion.
Verification
1. After calibration is complete and the detector is in normal opera-
tion, verify calibration by using a gas cylinder other than the one
used for calibration.
2. The gas concentration should not exceed the sensor’s detection
range. Confirm that the LCD displays the expected concentration
values.
3. To ensure that the reading is accurate, apply the verification gas
for the same period of time as was applied to the sensor when it
was calibrated.
Example:
SO
2
span time was 2 minutes therefore, apply verifica-
tion gas for 2 minutes.
Содержание GasAlert Extreme
Страница 1: ......
Страница 6: ...GasAlert Extreme User Manual ...
Страница 8: ...GasAlert Extreme User Manual ...
Страница 10: ...GasAlert Extreme User Manual ...
Страница 62: ...GasAlert Extreme User Manual ...
Страница 63: ...iERP 125465 D5561 6 English BW Technologies 2008 All rights reserved ...