27
GasAlert Extreme
User Manual
Table 6. Alarms
Note
The high alarm and STEL alarm have the same priority. A high alarm and/or STEL alarm overrides a low alarm and/or TWA alarm. To check
STEL and TWA alarms specifically, press and hold
C
and
D
simultaneously.
The vibrator alarm is disabled at -20°C.
The high and low alarms deactivate when the gas concentration is lower than the low alarm setpoint. If the alarms are set to latch, alarms
persist until the gas concentration is below the alarm setpoint and the alarms have been acknowledged by pressing
C
. The TWA and STEL
alarms deactivate by clearing the TWA and STEL peak exposure. Refer to
Clearing Gas Exposures
.
Alarm
Display
Alarm
Display
Automatic Shutdown Alarm:
(Low battery)
•
Eight beeps, flashes, and vibrations
•
LOW
displays
Automatic Shutdown Alarm:
(Calibration past)
•
Eight beeps, flashes, and vibrations
Automatic Shutdown Alarm:
(After Automatic Shutdown Alarm)
•
No beep
•
No flash or vibrations
•
displays for a short time
Confidence Beep:
•
One beep every 5 seconds
•
One quick vibration per minute
Note
If the detector enters low battery
alarm, the confidence beep deacti-
vates.