GasAlert Extreme
User Manual
11
9.
Low Alarm Setpoint:
The LCD displays the low alarm setpoint.
10.
High Alarm Setpoint:
The LCD displays the high alarm setpoint.
11.
Calibration Due Test:
The LCD displays the calibration due
date.
The LCD displays the number of days remaining before the
detector must be calibrated. For more information, refer to
Cali-
bration
.
If calibration is past due,
CAL. PAST
displays.
Press
C
to acknowledge the warning message.
If the calibration past due option is enabled or disabled, one of
the following two events will occur.
•
Calibration Past Due Disabled:
If the detector is not pass-
code protected and after the
CAL. PAST
message is acknowl-
edged, the detector continues the self-test and then enters
normal operation.
•
Calibration Past Due Enabled:
If the detector is passcode
protected and when
CAL. PAST
displays, press
C
to
acknowledge the message and to access the
PASS
screen. If
required, refer to
Passcode Protection Option
.
Press
D
or
E
to scroll to the required passcode, and press
C
within 10 seconds to confirm the selection. The detector
enters normal operation.
Note
Calibrate the detector before continuing operation.
If the passcode is not confirmed within 10 seconds or the pass-
code is incorrect, the detector beeps eight times, the LEDs
flash 8 times, and the LCD displays the following screen.