GasAlert Extreme
User Manual
21
Calibration Past Due Option
The calibration past due (
PAST
) option enables an automatic shutdown
during start-up if the detector is past due for calibration.
Note
The detector is shipped with the calibration past due shutdown
option disabled.
To calibrate a past due calibration detector, refer to
Calibration Past Due
Enabled
in
Self-Test
.
To enable/disable the calibration past due automatic shutdown option,
complete the following:
1. From the
EXIT
screen of the user options menu, press
D
or
E
to scroll to the
PAST
option.
2. Press
C
to select the option. The LCD flashes either
ON
or
OFF
.
3. Press
D
or
E
to toggle between the
ON
/
OFF
options. Ensure
the desired option is displayed and press
C
to confirm the
selection.
The LCD returns to the
EXIT
screen.
4. Press
D
or
E
to scroll to a new user option or press
C
to exit
and return to normal operation.
Note
If the option is not confirmed by pressing
C
within 10 seconds,
the detector returns to normal operation.
Languages
The LCD can display text in five different languages. Refer to the follow-
ing language options.
Portuguese Option
The Portuguese (
PORT
) option enables the LCD to display text in Portu-
guese.
Note
If the multi-language option is included, the detector is shipped
with English selected as the default language.
1. From the
EXIT
screen of the user options menu, press
D
or
E
to scroll to the
PORT
option.
2. Press
C
to select the option. The LCD then displays the Portu-
guese exit screen.
3. Press
D
or
E
to scroll to another user option or press
C
to
exit and return to normal operation.