Unsuccessful Calibration
The calibration due date cannot be set unless all sensors
successfully span.
If the calibration is unsuccessful, or if calibration is due but is
not performed, the following error message displays and
scrolls across the bottom of the LCD.
Verification
After calibration is complete and the detector is in normal
operating mode, test the detector by using a gas cylinder
other than the one used in calibration. The gas concentration
should not exceed the sensor's detection range. Verify that
the detector LCD displays the expected concentration.