GasAlertMicro
Calibration and Setting Alarm Setpoints
Calibration Procedures
The calibration process requires several functions, some of
which can be bypassed. A note is added to each option that
can be bypassed.
Note
Verify that the calibration gas being used matches
the span concentration value(s) that are defined for
the detector. For more information, refer to
.
The following procedures for calibration and setting alarm
setpoints are listed in the same order as they display on the
detector. They are as follows:
•
Start Calibration
•
Auto Zero and Oxygen (O
2
) Sensor Calibration
•
Pass Code Protect Activated
•
Auto Span
•
Alarm Setpoints
•
Setting the Calibration Due Date
Start Calibration
Note
To bypass calibration, at any time after auto zero
press
A
. The detector retains all saved values and
the audible alarm beeps four times before the
detector proceeds to the alarm setpoint screens.
To begin the calibrate/set alarm setpoints process, complete
the following:
1. Ensure the following procedures are performed in a
clean atmosphere.
2. From normal operating mode, press and hold
C
and
H
until the calibration screen displays.
43