6.
Calibration Status Test:
Calibration status displays the
number of days remaining before calibration is due.
If calibration is overdue, an audible alarm sounds and the
LCD displays the number of days the calibration is overdue.
To acknowledge the alarm, complete the following:
Press
C
to acknowledge and silence the audible alarm.
Once in normal operation mode, calibrate the detector. Refer
to
Calibration and Setting Alarm Points
.
7.
Bump Check Reminder Test
(optional): The detector
can be enabled to remind when a bump check is due. If this
feature is enabled and bump check is due, the LCD displays
the following screens:
The detector beeps twice, flashes once, and emits one
vibration.
1. To temporarily bypass the reminder, hold the
C
button for two seconds.
Note
Bypassing the reminder is temporary. The reminder
will display and alarm the next time the detector is
activated.
2. To bump the detector using the MicroDock II
Automatic Test and Calibration Station, refer to the
MicroDock ll User Manual.
If the detector has recently failed a bump test in the
MicroDock II, the detector’s LCD displays
nOtE: bumP tESt
hAS FAiLEd
.
3. Press
C
to acknowledge the warning.
4. From the MicroDock ll run another bump check.