92
MAN0604_Issue 09_02-2013
Apex
2110M8030
5. MAINTENANCE
Configuration Menu
Change Cartridge
Configure 4-20mA
Configure Relays
Configure Alarms
Configure Digital
Select Language
Configure Backlight
(5) Select
Change Cartridge
from the menu list.
(6) Press
ok
.
The
Change Cartridge
screen is displayed.
Change Cartridge
Change Cartridge Now
Cartridge: Present
Whilst this screen is showing, the cartridge can be safely replaced with the unit powered-up
as follows.
(7)
Remove the Certified Sensor cap or accessory from the sensor body.
Rotate the cap or accessory 1/4 turn in an anticlockwise direction to release the bayonet
fitting.
(8)
Remove the existing cartridge.
The cartridge plugs into two sockets in the Certified Sensor body and is located by means of
a tab.
To remove the cartridge without damaging it, grasp the visible part of the cartridge and pull it
firmly straight out of the sensor body without twisting. Slight resistance is initially felt as the
cartridge is pulled out of the electrical sockets.
Sensor
body
Sensor
cap