4.5 USER TASKS
The following table provides a quick reference list of typical tasks and how to carry them out via the
menu system.
Tasks related to:
Sub-Menu
Sub-Menu Option User Level
4-20mA loop
Set Inhibit, Warning and Overrange levels
Configuration
Configure 4-20mA
2
Zero setting
Calibration
4-20mA Set Zero
2
Test operation
Calibration
Force 4-20mA
2
Alarms
Set overrange latching on or off
Configuration
Configure Alarms
2
Set gas alarm levels
Configuration
Configure Alarms
2
Reset alarms
Display
Reset Alarms
1
Checking the unit’s settings
Calibration settings
Display
Calibration Information
1
Configuration settings
Display
Current Configuration
1
Checking the unit’s operational history
View the unit's log
History Log
View History Log
2
Clearing the history log
History Log
Reset History Log
2
Digital communications
View settings
Configuration Configure Digital
2
Register the communications board
Configuration Configure Digital
2
Test operation
Calibration Force Digital
2
Display Screen
Turn backlight on/off *
Configuration
Configure Backlight
2
Change the on-screen display language
Configuration
Select Language
2
gas
Test unit with gas
Calibration
Gas Challenge
2
Change a gas sensor cartridge
Configuration
Change Cartridge
2
Calibrate gas
Calibration
Gas Calibrate
2
Passwords
Change a password
Change Passwords Level 1 Password
3
Change Passwords Level 2 Password
3
Change Passwords Level 3 Password
3
Reset all passwords to default
Reset passwords
3
Relays
Set-up relay contacts
Configuration
Configure Relays
2
Test relays
Calibration
Force Relays
2
*
The
Configure Backlight
control is not available on CSA Certified Transmitter Units.