58
MAN0604_Issue 09_02-2013
Apex
2110M8030
4. OPERATION
Gas Calibrate
Calibration Menu
This option calibrates the Certified Sensor zero gas point and the span setting point (see
4.7
for
details of how to carry out this procedure)
.
Gas Calibrate
ok
Zero Gas
Zero Calibration screen
Calibrating
Success?
Yes
No
Span calibration?
ok
ok
Fault
diagnosis
Yes
No
Calibration
Menu
ok
Span Calibration screen
Set Span
Concentration
(ppm)
ok
Use up/
down
buttons
Apply Gas
ok
Calibrating
Success?
Yes
No
ok
Fault
diagnosis
Calibration Menu
Note: Depending on the type of gas cartridge fitted to the sensor, a choice of two calibration gases
may be displayed during
Span Calibration
,
e.g. C
7
H
8
has the options C
7
H
8
and CH
4
as choices. If so choose the one that will be used for
calibration and press
ok
to continue
.