Under these conditions you should close the external shut off valves:
•
The system is shut down for an extended period, taken out of service or
relocated.
•
The venturi pumps or gas cell are serviced.
•
The utility external connections (
Figure 25
) or the internal supply lines are
disconnected.
CAUTION!
This is a high pressure safety hazard. The compressed air (CDA) supply
pressure normally exceeds 100 psig (7 kg/cm2). After shutting off the external
supply valve, check the internal pressure regulators on the CDA Panel,
Figure
20
, to ensure that they read zero. This confirms that the pressure has bled
down and that it is safe to continue.
MISE EN GARDE!
Ceci est un risque de sécurité à haute pression . L'air comprimé ( CDA ) offre
pression dépasse normalement 100 psig ( 7 kg / cm2) . Après arrêt de l'
externe fournir vanne , vérifier les régulateurs de pression internes sur le
panneau de CDA , figure 20 , afin d'assurer qu'ils lisent zéro. Cela confirme
que la pression a saigné vers le bas et qu'il est sûr de continuer .
Internal disconnects
If you need to disconnect a specific assembly from line power to prevent it from
starting when power is restored, unplug it from the Power Distribution Panel,
Figure
19
.
Содержание ACM 150
Страница 1: ...ACM 150 Air Composition Monitor Guide to Installation Operation and Maintenance ...
Страница 2: ......
Страница 15: ...ACM 150 15 Figure 1 ACM 150 outside labels ...
Страница 16: ...ACM 150 16 Figure 2 ACM 150 inside labels ...
Страница 29: ...ACM 150 29 Figure 15 Relay output board ...
Страница 59: ...ACM 150 59 Figure 30 Installation diagram ...
Страница 152: ...ACM 150 152 ...
Страница 153: ......
Страница 154: ......
Страница 155: ......
Страница 156: ...ACM 150 Air Composition Monitor Revision 2 ECO HAA150063 December 2015 Honeywell Analytics ...