ACM 150
136
Resolving problems
Table 6
lists service messages that might be generated by the ACM 150 monitor
software. Troubleshoot the problems by following the recommended actions. Use
‘refer to’ for guidance, additional information and/or step-by-step procedures.
Table 6. Service request messages
Service Message
Recommended Action
Refer To
Possible line leak prob-
lems in line xx
Test the line and its
check valve for leaks
•
Line leak tests
(optional)
Possible line leak prob-
lems in lines (all)
Check venturi regulator
& air supply Check the
line leak setup
•
CDA panel
•
Line leak test setup
procedure
Energy (ZPD) value too
low (or too high)
Evaluate the optical
path
•
Evaluating the FTIR
optical path
No data from bench
Check the bench for a
malfunction Evaluate
optical path
•
Evaluating the FTIR
optical path
System stopped
No action required
Stopped at user request
System started
No action required
Stopped at user request
No active scans are
defined
Set up an Order of Scan
•
Order of scan
Cannot establish com-
munication with control
boards
Verify that the cables
and power cords are
connected to the con-
trol boards
•
Interconnecting
cables
Cannot update relays
Verify that the cables
and power cords are
connected to the relay
board(s)
•
Interconnecting
cables
Cannot load factory
background spectra
Check for missing file
or path error (should not
occur after initial start-
up)
•
Contact Honeywell
Analytics
Energy (ZPD) address
out of range or can’t
determine ZPD
Too little reaching de-
tector
•
Evaluating the FTIR
optical path
Scan cancelled
Normal operation
Incorrect data length
(short/long scan)
Poor connection at FTIR
data cable
•
Evaluating the FTIR
optical path
Can not evaluate equa-
tion for relay XX
SQL error in relay setup,
advanced tab
•
Virtual relay setup.
Содержание ACM 150
Страница 1: ...ACM 150 Air Composition Monitor Guide to Installation Operation and Maintenance ...
Страница 2: ......
Страница 15: ...ACM 150 15 Figure 1 ACM 150 outside labels ...
Страница 16: ...ACM 150 16 Figure 2 ACM 150 inside labels ...
Страница 29: ...ACM 150 29 Figure 15 Relay output board ...
Страница 59: ...ACM 150 59 Figure 30 Installation diagram ...
Страница 152: ...ACM 150 152 ...
Страница 153: ......
Страница 154: ......
Страница 155: ......
Страница 156: ...ACM 150 Air Composition Monitor Revision 2 ECO HAA150063 December 2015 Honeywell Analytics ...