ACM 150
52
Sample area connections
NOTE
The buyer is responsible for deciding which areas and gases to monitor and
the exact location of the sampling points in gas cabinets, storage rooms,
valve boxes, exhaust ducts, exhaust hoods, breathing zones, etc. Honeywell
Analytics does not determine and cannot advise the exact point to sample
for hazardous gases. The buyer, the equipment manufacturer or a third party
consultant must make such decisions.
Figure 26. ACM 150 Tube connection points
Points and gases monitored
The ACM 150 monitor is designed for and intended to detect occasional leaks
of hazardous gases at levels that are within its range of detection. The following
conditions may harm the ACM 150 monitor and may violate the warranty terms:
•
Continuous presence of oxidizing or corrosive gases at one or more sample point
at levels exceeding 10 PPM, particularly if the air is humid.
•
High PPM level leaks that saturate the ACM 150 monitor with oxidizing or
corrosive vapors.
•
Liquids (water, solvents or oil) present in the sampled air that enter the ACM 150
monitor via the sample tubing.
•
Flammable gases above explosive levels that might increase the risk of fire inside
the ACM 150 monitor.
The ACM 150 monitor is not suited for monitoring samples that contain high levels of
corrosive and/ or condensable gases, such as the exhaust from treatment systems.
Application suitability and adaptability
Honeywell Analytics provides a list of typical gases for the ACM 150 monitor and
Содержание ACM 150
Страница 1: ...ACM 150 Air Composition Monitor Guide to Installation Operation and Maintenance ...
Страница 2: ......
Страница 15: ...ACM 150 15 Figure 1 ACM 150 outside labels ...
Страница 16: ...ACM 150 16 Figure 2 ACM 150 inside labels ...
Страница 29: ...ACM 150 29 Figure 15 Relay output board ...
Страница 59: ...ACM 150 59 Figure 30 Installation diagram ...
Страница 152: ...ACM 150 152 ...
Страница 153: ......
Страница 154: ......
Страница 155: ......
Страница 156: ...ACM 150 Air Composition Monitor Revision 2 ECO HAA150063 December 2015 Honeywell Analytics ...