Figure 67. Background, S/N, and drift
Acquire background
You can obtain a background spectrum on demand by selecting ‘Acquire
Background’. As shown in
Figure 68
, the background spectrum is displayed for
evaluation.
In this mode it is also possible to ‘Save New Factory Background’ Scan’, which
would replace the prior reference background.
NOTE
Do not save a new factory background unless you are requested to do so by
your HA service technician or application chemist. There should be a valid
purpose for updating this reference.