HA7062D Manual Rev 1.01
www.holzworth.com
Page 77 of 105
Frequency Offset Start/Stop
:SENS:PN:FREQ:STAR:<value>
Sets frequency offset start (how close to the
carrier to measure)
Example
TX:
:SENS:PN:FREQ:STAR:100Hz
RX:
Frequency start set
:SENS:PN:FREQ:STAR?
Query measurement start frequency
Example
TX:
:SENS:PN:FREQ:STAR?
RX:
100
:SENS:PN:FREQ:STOP:<value>
Sets frequency offset stop (how far from the
carrier to measure)
Example
TX:
:SENS:PN:FREQ:STOP:10MHz
RX:
Frequency stop set
:SENS:PN:FREQ:STOP?
Query measurement stop frequency
Example
TX:
:SENS:PN:FREQ:STOP?
RX:
10000000