HA7062D Manual Rev 1.01
www.holzworth.com
Page 66 of 105
3.
In the Frequency Span and Trigger/Averaging/Bandwidth sub-sections, adjust measurement
settings as necessary for the DUT including measurement offset range, limits for integration to
calculate jitter, data resolution, and the number of cross correlations to be performed.
4.
Click the Inputs button on the right side of the software interface. Select HX5100 Phase Shifters
in the DUT Inputs sub-section of the Inputs menu.
Also in the Inputs
menu, verify the LO power levels by clicking the Measure Input Frequency and
Power
button in the
LO Test
section. LO power levels should be in the range of
+7 to +13 dBm
for
optimal measurement sensitivity.
5.
When hardware and software configuration is complete, and LO power levels are within range,
click the
'Acquire'
button to begin the measurement.
The correlation count will be displayed in the bottom of the plot area, and the measurement time
remaining is displayed at the bottom left of the screen by the Holzworth shockwave. The
shockwave also acts as a progress bar.