HA7062D Manual Rev 1.01
www.holzworth.com
Page 8 of 105
6.1.2 FRONT PANEL CONNECTORS
DESCRIPTION
SPECIFICATION
Front Panel Connector(s) Type
26GHz DUT Input (standard)
40GHz DUT Input (optional)
All other
3.5mm (50ohm)
2.92mm (50ohm)
3.5mm (50ohm)
DUT Input
26GHz Frequency Range (standard)
40GHz Frequency Range (optional)
Power Level Range
Input Damage Level
10MHz to 26GHz
10MHz to 40GHz
-5dBm to +20dBm
> +22dBm, >25Vdc
DUT Tune Voltage
Voltage Tune Range
Max Current
For Vcc control of some DUTs.
-10V to +12V
5mA
DUT Power Supply
Voltage Supply Range
Maximum Current
Integrated power supply.
0V to +12V
250mA
LO1/LO2 Input(s)
Frequency Range (standard)
Power Level Range
Input Damage Level
Connect to
LO1/LO2 Output(s)
for standard operation.
10MHz to 6GHz
+3dBm to +13dBm
> +22dBm
LO1/LO2 Output(s)
Frequency Range (standard)
Power Level Range
Connect to
LO1/LO2 Input(s)
for standard operation.
10MHz to 6GHz (0.001Hz step size)
0dBm to +10dBm (0.01dB step size)
LO1/LO2 Tune Voltage
Voltage Tune Range
Max Current
Tuning Sensitivity
For independent Vcc control of external LO sources during External LO Mode of operation.
-10V to +12V
5mA
TBD
DUT ch1/ch2 Input(s)
Frequency Range (standard)
Power Level Range
Input Damage Level
Allows bypass of DUT power splitter for direct access to the phase detector of each channel.
10MHz to 6GHz
0dBm to +14dBm
> +16dBm
ch1/ch2 Splitter Output(s)
Connect to DUT ch1/ch2 Inputs for standard operation.
ch1/ch2 Baseband Input(s)
Frequency Range
Power Level Range
Input Damage Level
DC to 100MHz
± 1Vdc
± 2Vdc, or 50mA (whichever is greater), +10dBm RF power