HA7062D Manual Rev 1.01
www.holzworth.com
Page 71 of 105
d.
Adjust the phase shifter so that channel 1 of the
DC Monitor
is reading 0.01000 V
(10mV). Again record the value displayed on the mechanical phase shifter.
In this example the reading is 54.44.
e.
Adjust the phase shifter so that channel 1 is back to as close to 0.00000V as possible.
Click
Stop
in the
DC Monitor
window.
f.
Calculate the delta in the phase shifter readings.
55.06 - 54.44 = 0.62
g.
Multiply the delta by the carrier frequency in GHz to calculate degrees.
0.62 x 3GHz = 1.86 degrees
h.
Convert degrees to radians and then divide volts/radian, where volts = 10mV.
1.86 degrees = 0.0325 radians
Kd = 10mV/0.0325 radians
Kd = 0.308
Enter the calculated value for
Kd: