HA7062D Manual Rev 1.01
www.holzworth.com
Page 12 of 105
7.3 MEASUREMENT SPEED (SAMPLE TIME) vs. MINIMUM OFFSET
NOTE:
Measurement times shown below are total data acquisition time. Data acquisition time does
not include data transfer time to PC. Utilizing an Ethernet/LAN connection is recommended to best
take advantage of the data acquisition speed of the instrument.
1 Cross-Correlation
64 Samples
128 Samples
256 Samples
512 Samples
1024 Samples
-
-
-
-
0.1Hz
17s
0.1Hz
34s
0.1Hz
1m8s
1Hz
1.1s
1Hz
2.2s
1Hz
4.3s
1Hz
8.6s
1Hz
17s
10Hz
0.3s
10Hz
0.5s
10Hz
1.1s
10Hz
2.2s
10Hz
4.3s
100Hz
0.1s
100Hz
0.1s
100Hz
0.3s
100Hz
0.5s
100Hz
1.1s
1kHz
<0.1s
1kHz
<0.1s
1kHz
0.1s
1kHz
0.1s
1kHz
0.3s
10kHz
<0.1s
10kHz
<0.1s
10kHz
<0.1s
10kHz
<0.1s
10kHz
<0.1s
100kHz
<0.1s
100kHz
<0.1s
100kHz
<0.1s
100kHz
<0.1s
100kHz
<0.1s
1MHz
<0.1s
1MHz
<0.1s
1MHz
<0.1s
1MHz
<0.1s
1MHz
<0.1s
10 Cross-Correlations
64 Samples
128 Samples
256 Samples
512 Samples
1024 Samples
-
-
-
-
0.1Hz
2m51s
0.1Hz
5m42s
0.1Hz
11m24s
1Hz
11s
1Hz
21s
1Hz
43s
1Hz
1m25s
1Hz
2m51s
10Hz
2.7s
10Hz
5.4s
10Hz
11s
10Hz
22s
10Hz
43s
100Hz
0.7s
100Hz
1.3s
100Hz
2.7s
100Hz
5.4s
100Hz
11s
1kHz
0.2s
1kHz
0.3s
1kHz
0.7s
1kHz
1.3s
1kHz
2.7s
10kHz
<0.1s
10kHz
<0.1s
10kHz
0.1s
10kHz
0.2s
10kHz
0.3s
100kHz
<0.1s
100kHz
<0.1s
100kHz
<0.1s
100kHz
<0.1s
100kHz
<0.1s
1MHz
<0.1s
1MHz
<0.1s
1MHz
<0.1s
1MHz
<0.1s
1MHz
<0.1s
7.4 xCORRELATIONS vs. PHASE NOISE IMPROVEMENT
Improvement factor: dB = 5logN (N = No. of correlations)
Number of Correlations
1
10
100
1,000
10,000
dB Improvement
0dB
5dB
10dB
15dB
20dB